High Sensitive ELISA Kit for Melatonin (MT)
N-Acetyl-5-Methoxytryptamine
- Product No.HEA908Ge
- Organism SpeciesPan-species (General) Same name, Different species.
- Test MethodCompetitive Inhibition
- Assay Length3h
- Detection Range7.8-500pg/mL
- SensitivityThe minimum detectable dose of this kit is typically less than 3.1pg/mL.
- Sample Typeserum, plasma and other biological fluids
- Download Instruction Manual
- UOM 48T96T 96T*5 96T*10 96T*100
-
FOB
US$ 613
For more details, please contact local distributors!US$ 876
For more details, please contact local distributors! US$ 3942
For more details, please contact local distributors! US$ 7446
For more details, please contact local distributors! US$ 61320
For more details, please contact local distributors!
Specificity of the High Sensitive ELISA Kit for Melatonin (MT)
This assay has high sensitivity and excellent specificity for detection of High Sensitive Melatonin (MT).
No significant cross-reactivity or interference between High Sensitive Melatonin (MT) and analogues was observed.
Recovery of the High Sensitive ELISA Kit for Melatonin (MT)
Matrices listed below were spiked with certain level of High Sensitive Melatonin (MT) and the recovery rates were calculated by comparing the measured value to the expected amount of High Sensitive Melatonin (MT) in samples.
Matrix | Recovery range (%) | Average(%) |
serum(n=5) | 96-105 | 101 |
EDTA plasma(n=5) | 88-102 | 96 |
heparin plasma(n=5) | 80-94 | 88 |
Precision of the High Sensitive ELISA Kit for Melatonin (MT)
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level High Sensitive Melatonin (MT) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level High Sensitive Melatonin (MT) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Linearity of the High Sensitive ELISA Kit for Melatonin (MT)
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of High Sensitive Melatonin (MT) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample | 1:2 | 1:4 | 1:8 | 1:16 |
serum(n=5) | 86-98% | 78-88% | 78-102% | 86-101% |
EDTA plasma(n=5) | 97-104% | 79-90% | 85-92% | 83-96% |
heparin plasma(n=5) | 78-92% | 96-104% | 87-95% | 80-102% |
Stability of the High Sensitive ELISA Kit for Melatonin (MT)
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Assay procedure summary of the High Sensitive ELISA Kit for Melatonin (MT)
1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
And then add 50µL prepared Detection Reagent A immediately.
Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
7. Add 50µL Stop Solution. Read at 450 nm immediately.
Test principle of the High Sensitive ELISA Kit for Melatonin (MT)
This assay employs the competitive inhibition enzyme immunoassay technique. A monoclonal antibody specific to High Sensitive Melatonin (MT) has been pre-coated onto a microplate. A competitive inhibition reaction is launched between biotin labeled High Sensitive Melatonin (MT) and unlabeled High Sensitive Melatonin (MT) (Standards or samples) with the pre-coated antibody specific to High Sensitive Melatonin (MT). After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of High Sensitive Melatonin (MT) in the sample. After addition of the substrate solution, the intensity of color developed is reverse proportional to the concentration of High Sensitive Melatonin (MT) in the sample.
GIVEAWAYS
INCREMENT SERVICES
Related products
Catalog No. | Organism species: Pan-species (General) | Applications (RESEARCH USE ONLY!) |
CPA908Ge11 | BSA Conjugated Melatonin (MT) | Immunogen; SDS-PAGE; WB. |
CPA908Ge21 | OVA Conjugated Melatonin (MT) | Immunogen; SDS-PAGE; WB. |
PAA908Ge01 | Polyclonal Antibody to Melatonin (MT) | ELISA, CLIA. / IHC-Fr, ICC, IP (predicted). |
MEA908Ge | Mini Samples ELISA Kit for Melatonin (MT) | Enzyme-linked immunosorbent assay for Antigen Detection. |
CEA908Ge | ELISA Kit for Melatonin (MT) | Enzyme-linked immunosorbent assay for Antigen Detection. |
HEA908Ge | High Sensitive ELISA Kit for Melatonin (MT) | Enzyme-linked immunosorbent assay for Antigen Detection. |
IEA908Ge | Instant ELISA Kit for Melatonin (MT) | Enzyme-linked immunosorbent assay for Antigen Detection. |
CCA908Ge | CLIA Kit for Melatonin (MT) | Chemiluminescent immunoassay for Antigen Detection. |
LMA908Ge | Multiplex Assay Kit for Melatonin (MT) ,etc. by FLIA (Flow Luminescence Immunoassay) | FLIA Kit for Antigen Detection. |
KSA908Ge11 | ELISA Kit DIY Materials for Melatonin (MT) | Main materials for "Do It (ELISA Kit) Yourself". |