Multiplex Assay Kit for Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay)

HAVCR1; TIM1; TIMD1; HAVCR; Hepatitis A Virus Cellular Receptor 1; T Cell Immunoglobulin And Mucin Domain-Containing Protein 1; T-cell immunoglobulin mucin receptor 1

(Note: Up to 8-plex in one testing reaction)

  • Multiplex Assay Kit for Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay) Packages (Simulation)
  • Multiplex Assay Kit for Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay) Packages (Simulation)
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the Multiplex Assay Kit for Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay)

This assay has high sensitivity and excellent specificity for detection of Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery of the Multiplex Assay Kit for Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay)

Matrices listed below were spiked with certain level of recombinant Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 94-101 97
EDTA plasma(n=5) 91-101 95
heparin plasma(n=5) 85-99 93

Precision of the Multiplex Assay Kit for Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the Multiplex Assay Kit for Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 79-101% 80-95% 83-101% 80-103%
EDTA plasma(n=5) 91-99% 88-104% 86-101% 87-103%
heparin plasma(n=5) 89-98% 81-99% 92-99% 91-98%

Stability of the Multiplex Assay Kit for Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the Multiplex Assay Kit for Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay)

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

Test principle of the Multiplex Assay Kit for Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay)

Analyte-specific antibodies are pre-coated onto color-coded microparticles. Microparticles, standards, and samples are pipetted into wells and the immobilized antibodies bind the analytes of interest. After washing away any unbound substances, a biotinylated antibody cocktail specific to the analytes of interest is added to each well. Following a wash to remove any unbound biotinylated antibody, Streptavidin-Phycoerythrin conjugate (Streptavidin-PE), which binds to the biotinylated detection antibodies, is added to each well. A final wash removes unbound Streptavidin-PE and the microparticles are resuspended in buffer and read using the Luminex or Bio-Plex analyzer.The MFI developed is proportional to the concentration of analytes of interest in the sample.

GIVEAWAYS

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
RPA785Hu01 Recombinant Kidney Injury Molecule 1 (Kim1) Positive Control; Immunogen; SDS-PAGE; WB.
RPA785Hu02 Recombinant Kidney Injury Molecule 1 (Kim1) Positive Control; Immunogen; SDS-PAGE; WB.
RPA785Hu03 Recombinant Kidney Injury Molecule 1 (Kim1) Positive Control; Immunogen; SDS-PAGE; WB.
PAA785Hu01 Polyclonal Antibody to Kidney Injury Molecule 1 (Kim1) WB; IHC; ICC; IP.
PAA785Hu02 Polyclonal Antibody to Kidney Injury Molecule 1 (Kim1) WB; IHC; ICC; IP.
PAA785Hu03 Polyclonal Antibody to Kidney Injury Molecule 1 (Kim1) WB; IHC; ICC; IP.
LAA785Hu81 FITC-Linked Polyclonal Antibody to Kidney Injury Molecule 1 (Kim1) WB; IHC; ICC; IF.
LAA785Hu71 Biotin-Linked Polyclonal Antibody to Kidney Injury Molecule 1 (Kim1) WB; IHC; ICC.
MAA785Hu22 Monoclonal Antibody to Kidney Injury Molecule 1 (Kim1) WB
MAA785Hu21 Monoclonal Antibody to Kidney Injury Molecule 1 (Kim1) WB; IHC; ICC; IP.
MAA785Hu23 Monoclonal Antibody to Kidney Injury Molecule 1 (Kim1) IHC
MAA785Hu24 Monoclonal Antibody to Kidney Injury Molecule 1 (Kim1) WB; IHC; ICC; IP.
SEA785Hu ELISA Kit for Kidney Injury Molecule 1 (Kim1) Enzyme-linked immunosorbent assay for Antigen Detection.
SCA785Hu CLIA Kit for Kidney Injury Molecule 1 (Kim1) Chemiluminescent immunoassay for Antigen Detection.
LMA785Hu Multiplex Assay Kit for Kidney Injury Molecule 1 (Kim1) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
PSA785Hu01 Antibody Pair for Kidney Injury Molecule 1 (Kim1) ELISA; CLIA; ELISPOT; Luminex; Immunochromatography and other Immunoassays.
KSA785Hu01 ELISA Kit DIY Materials for Kidney Injury Molecule 1 (Kim1) Main materials for "Do It (ELISA Kit) Yourself".
GUESTBOOK