Mini Samples ELISA Kit for Glutathione (GSH)
L-Reduced Glutathione
- Product No.MEA294Ge
- Organism SpeciesPan-species (General) Same name, Different species.
- Test MethodCompetitive Inhibition
- Assay Length3h
- Detection Range1.23-100ug/mL
- SensitivityThe minimum detectable dose of this kit is typically less than 0.51ug/mL.
- Sample Typeserum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
- Download Instruction Manual
- UOM 48T96T 96T*5 96T*10 96T*100
-
FOB
US$
For more details, please contact local distributors!US$
For more details, please contact local distributors! US$
For more details, please contact local distributors! US$
For more details, please contact local distributors! US$
For more details, please contact local distributors!
Specificity of the Mini Samples ELISA Kit for Glutathione (GSH)
This assay has high sensitivity and excellent specificity for detection of Mini Samples Glutathione (GSH).
No significant cross-reactivity or interference between Mini Samples Glutathione (GSH) and analogues was observed.
Recovery of the Mini Samples ELISA Kit for Glutathione (GSH)
Matrices listed below were spiked with certain level of Mini Samples Glutathione (GSH) and the recovery rates were calculated by comparing the measured value to the expected amount of Mini Samples Glutathione (GSH) in samples.
Matrix | Recovery range (%) | Average(%) |
serum(n=5) | 94-101 | 97 |
EDTA plasma(n=5) | 78-104 | 86 |
heparin plasma(n=5) | 79-97 | 88 |
Precision of the Mini Samples ELISA Kit for Glutathione (GSH)
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Mini Samples Glutathione (GSH) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Mini Samples Glutathione (GSH) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Linearity of the Mini Samples ELISA Kit for Glutathione (GSH)
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Mini Samples Glutathione (GSH) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample | 1:2 | 1:4 | 1:8 | 1:16 |
serum(n=5) | 83-101% | 99-105% | 83-98% | 89-104% |
EDTA plasma(n=5) | 90-98% | 87-105% | 94-102% | 80-99% |
heparin plasma(n=5) | 83-101% | 98-105% | 83-101% | 78-90% |
Stability of the Mini Samples ELISA Kit for Glutathione (GSH)
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Assay procedure summary of the Mini Samples ELISA Kit for Glutathione (GSH)
1. Prepare all reagents, samples and standards;
2. Add 25µL standard or sample to each well.
And then add 25μL prepared Detection Reagent A immediately.
Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 50µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 50µL Substrate Solution. Incubate 10-20 minutes at 37°C;
7. Add 25µL Stop Solution. Read at 450 nm immediately.
Test principle of the Mini Samples ELISA Kit for Glutathione (GSH)
This assay employs the competitive inhibition enzyme immunoassay technique. A monoclonal antibody specific to Mini Samples Glutathione (GSH) has been pre-coated onto a microplate. A competitive inhibition reaction is launched between biotin labeled Mini Samples Glutathione (GSH) and unlabeled Mini Samples Glutathione (GSH) (Standards or samples) with the pre-coated antibody specific to Mini Samples Glutathione (GSH). After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of Mini Samples Glutathione (GSH) in the sample. After addition of the substrate solution, the intensity of color developed is reverse proportional to the concentration of Mini Samples Glutathione (GSH) in the sample.
GIVEAWAYS
INCREMENT SERVICES
Related products
Catalog No. | Organism species: Pan-species (General) | Applications (RESEARCH USE ONLY!) |
SPA294Ge01 | Synthetic Glutathione (GSH) | Immunogen; Blocking Peptide. |
CPA294Ge11 | BSA Conjugated Glutathione (GSH) | Immunogen; SDS-PAGE; WB. |
CPA294Ge21 | OVA Conjugated Glutathione (GSH) | Immunogen; SDS-PAGE; WB. |
CPA294Ge41 | HSA Conjugated Glutathione (GSH) | Immunogen; SDS-PAGE; WB. |
PAA294Ge01 | Polyclonal Antibody to Glutathione (GSH) | ELISA, CLIA. / IHC-Fr, ICC, IP (predicted). |
LAA294Ge71 | Biotin-Linked Polyclonal Antibody to Glutathione (GSH) | WB; IHC; ICC. |
MAA294Ge22 | Monoclonal Antibody to Glutathione (GSH) | ELISA, CLIA. / IHC-Fr, ICC, IP (predicted). |
MAA294Ge21 | Monoclonal Antibody to Glutathione (GSH) | ELISA, CLIA. / IHC-Fr, ICC, IP (predicted). |
CEA294Ge | ELISA Kit for Glutathione (GSH) | Enzyme-linked immunosorbent assay for Antigen Detection. |
MEA294Ge | Mini Samples ELISA Kit for Glutathione (GSH) | Enzyme-linked immunosorbent assay for Antigen Detection. |
LMA294Ge | Multiplex Assay Kit for Glutathione (GSH) ,etc. by FLIA (Flow Luminescence Immunoassay) | FLIA Kit for Antigen Detection. |
KSA294Ge11 | ELISA Kit DIY Materials for Glutathione (GSH) | Main materials for "Do It (ELISA Kit) Yourself". |