Mini Samples ELISA Kit for Aspartate Aminotransferase (AST)
cCAT; AAT; ASAT; SGOT; GOT1; Cysteine transaminase, cytoplasmic; Transaminase A; Aspartate Transaminase 1,Cytoplasmic; Glutamic-Oxaloacetic Transaminase 1,Soluble
- Product No.MEB214Hu
- Organism SpeciesHomo sapiens (Human) Same name, Different species.
- Test MethodDouble-antibody Sandwich
- Assay Length3h
- Detection Range15.6-1,000ng/mL
- SensitivityThe minimum detectable dose of this kit is typically less than 5.5ng/mL.
- Sample Typeserum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
- Download Instruction Manual
- UOM 48T96T 96T*5 96T*10 96T*100
-
FOB
US$ 559
For more details, please contact local distributors!US$ 798
For more details, please contact local distributors! US$ 3591
For more details, please contact local distributors! US$ 6783
For more details, please contact local distributors! US$ 55860
For more details, please contact local distributors!
Specificity of the Mini Samples ELISA Kit for Aspartate Aminotransferase (AST)
This assay has high sensitivity and excellent specificity for detection of Mini Samples Aspartate Aminotransferase (AST).
No significant cross-reactivity or interference between Mini Samples Aspartate Aminotransferase (AST) and analogues was observed.
Recovery of the Mini Samples ELISA Kit for Aspartate Aminotransferase (AST)
Matrices listed below were spiked with certain level of recombinant Mini Samples Aspartate Aminotransferase (AST) and the recovery rates were calculated by comparing the measured value to the expected amount of Mini Samples Aspartate Aminotransferase (AST) in samples.
Matrix | Recovery range (%) | Average(%) |
serum(n=5) | 80-88 | 80 |
EDTA plasma(n=5) | 84-98 | 94 |
heparin plasma(n=5) | 85-97 | 89 |
Precision of the Mini Samples ELISA Kit for Aspartate Aminotransferase (AST)
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Mini Samples Aspartate Aminotransferase (AST) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Mini Samples Aspartate Aminotransferase (AST) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Linearity of the Mini Samples ELISA Kit for Aspartate Aminotransferase (AST)
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Mini Samples Aspartate Aminotransferase (AST) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample | 1:2 | 1:4 | 1:8 | 1:16 |
serum(n=5) | 97-105% | 80-96% | 80-90% | 96-103% |
EDTA plasma(n=5) | 84-97% | 98-105% | 95-103% | 86-101% |
heparin plasma(n=5) | 87-101% | 95-103% | 87-101% | 98-105% |
Stability of the Mini Samples ELISA Kit for Aspartate Aminotransferase (AST)
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Assay procedure summary of the Mini Samples ELISA Kit for Aspartate Aminotransferase (AST)
1. Prepare all reagents, samples and standards;
2. Add 25µL standard or sample to each well. Incubate 1 hour at 37°C;
3. Aspirate and add 25µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 25µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 25µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 20µL Stop Solution. Read at 450nm immediately.
Test principle of the Mini Samples ELISA Kit for Aspartate Aminotransferase (AST)
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mini Samples Aspartate Aminotransferase (AST). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Mini Samples Aspartate Aminotransferase (AST). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mini Samples Aspartate Aminotransferase (AST), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mini Samples Aspartate Aminotransferase (AST) in the samples is then determined by comparing the O.D. of the samples to the standard curve.