ELISA Kit for Phosphodiesterase 12 (PDE12)
2'-PDE; 2'-Phosphodiesterase; Mitochondrial deadenylase
- Product No.USEF658Mu
- Organism SpeciesMus musculus (Mouse) Same name, Different species.
- Test MethodDouble-antibody Sandwich
- Assay Length3h
- Detection Range0.156-10ng/mL
- SensitivityThe minimum detectable dose of this kit is typically less than 0.051ng/mL.
- Sample TypeSerum, plasma, tissue homogenates and other biological fluids.
- Download Instruction Manual
- UOM 48T96T 96T*5 96T*10 96T*100
- 
    FOB
    
                        US$ 252 
 For more details, please contact local distributors!US$ 360
 For more details, please contact local distributors! US$ 1620
 For more details, please contact local distributors! US$ 3060
 For more details, please contact local distributors! US$ 25200
 For more details, please contact local distributors!
Specificity of the ELISA Kit for Phosphodiesterase 12 (PDE12)
                        This assay has high sensitivity and excellent specificity for detection of Phosphodiesterase 12 (PDE12).
                        No significant cross-reactivity or interference between Phosphodiesterase 12 (PDE12) and analogues was observed.
                    
Recovery of the ELISA Kit for Phosphodiesterase 12 (PDE12)
Matrices listed below were spiked with certain level of recombinant Phosphodiesterase 12 (PDE12) and the recovery rates were calculated by comparing the measured value to the expected amount of Phosphodiesterase 12 (PDE12) in samples.
| Matrix | Recovery range (%) | Average(%) | 
| serum(n=5) | 84-96 | 87 | 
| EDTA plasma(n=5) | 89-96 | 92 | 
| heparin plasma(n=5) | 95-104 | 101 | 
Precision of the ELISA Kit for Phosphodiesterase 12 (PDE12)
                    Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Phosphodiesterase 12 (PDE12) were tested 20 times on one plate, respectively. 
                    Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Phosphodiesterase 12 (PDE12) were tested on 3 different plates, 8 replicates in each plate. 
                    CV(%) = SD/meanX100 
                    Intra-Assay: CV<10% 
                    Inter-Assay: CV<12% 
                
Linearity of the ELISA Kit for Phosphodiesterase 12 (PDE12)
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Phosphodiesterase 12 (PDE12) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Sample | 1:2 | 1:4 | 1:8 | 1:16 | 
| serum(n=5) | 87-98% | 91-102% | 90-98% | 78-103% | 
| EDTA plasma(n=5) | 79-88% | 80-91% | 81-102% | 91-101% | 
| heparin plasma(n=5) | 82-94% | 78-103% | 89-103% | 83-99% | 
Stability of the ELISA Kit for Phosphodiesterase 12 (PDE12)
                    The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. 
                    To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
                
Assay procedure summary of the ELISA Kit for Phosphodiesterase 12 (PDE12)
                    1. Prepare all reagents, samples and standards;
                            2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
                            3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
                            4. Aspirate and wash 3 times;
                            5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
                            6. Aspirate and wash 5 times;
                            7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
                            8. Add 50µL Stop Solution. Read at 450nm immediately. 
                
Test principle of the ELISA Kit for Phosphodiesterase 12 (PDE12)
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Phosphodiesterase 12 (PDE12). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Phosphodiesterase 12 (PDE12). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Phosphodiesterase 12 (PDE12), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Phosphodiesterase 12 (PDE12) in the samples is then determined by comparing the O.D. of the samples to the standard curve.
GIVEAWAYS
INCREMENT SERVICES
                    Single-component Reagents of Assay Kit
                            Lysis Buffer Specific for ELISA / CLIA
                            Quality Control of Kit
                            ELISA Kit Customized Service
                            Disease Model Customized Service
                            Serums Customized Service
                            TGFB1 Activation Reagent
                            Real Time PCR Experimental Service
                            Streptavidin
                            Fast blue Protein Stain solution
                            Single-component Reagents of FLIA Kit
                            Streptavidin-Agarose Beads
                                            
Related products
| Catalog No. | Organism species: Mus musculus (Mouse) | Applications (RESEARCH USE ONLY!) | 
| URPF658Mu02 | Recombinant Phosphodiesterase 12 (PDE12) | Positive Control; Immunogen; SDS-PAGE; WB. | 
| URPF658Mu01 | Recombinant Phosphodiesterase 12 (PDE12) | Positive Control; Immunogen; SDS-PAGE; WB. | 
| UPAF658Mu01 | Polyclonal Antibody to Phosphodiesterase 12 (PDE12) | IHC | 
| UPAF658Mu02 | Polyclonal Antibody to Phosphodiesterase 12 (PDE12) | WB; IHC; ICC; IP. | 
| USEF658Mu | ELISA Kit for Phosphodiesterase 12 (PDE12) | Enzyme-linked immunosorbent assay for Antigen Detection. | 
| ULMF658Mu | Multiplex Assay Kit for Phosphodiesterase 12 (PDE12) ,etc. by FLIA (Flow Luminescence Immunoassay) | FLIA Kit for Antigen Detection. | 

 
                     
                    