ELISA Kit for Leptin (LEP)
OB; OBS; Obesity Homolog; Obesity Factor; Obese Protein
- Product No.SEA084Ov
- Organism SpeciesOvis aries; Ovine (Sheep) Same name, Different species.
- Test MethodDouble-antibody Sandwich
- Assay Length3h
- Detection Range0.156-10ng/mL
- SensitivityThe minimum detectable dose of this kit is typically less than 0.061ng/mL.
- Sample Typeserum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
- Download Instruction Manual
- UOM 48T96T 96T*5 96T*10 96T*100
-
FOB
US$ 567
For more details, please contact local distributors!US$ 810
For more details, please contact local distributors! US$ 3645
For more details, please contact local distributors! US$ 6885
For more details, please contact local distributors! US$ 56700
For more details, please contact local distributors!
Specificity of the ELISA Kit for Leptin (LEP)
This assay has high sensitivity and excellent specificity for detection of Leptin (LEP).
No significant cross-reactivity or interference between Leptin (LEP) and analogues was observed.
Recovery of the ELISA Kit for Leptin (LEP)
Matrices listed below were spiked with certain level of recombinant Leptin (LEP) and the recovery rates were calculated by comparing the measured value to the expected amount of Leptin (LEP) in samples.
Matrix | Recovery range (%) | Average(%) |
serum(n=5) | 98-105 | 101 |
EDTA plasma(n=5) | 78-101 | 84 |
heparin plasma(n=5) | 97-105 | 102 |
Precision of the ELISA Kit for Leptin (LEP)
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Leptin (LEP) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Leptin (LEP) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Linearity of the ELISA Kit for Leptin (LEP)
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Leptin (LEP) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Sample | 1:2 | 1:4 | 1:8 | 1:16 |
serum(n=5) | 86-94% | 92-105% | 87-101% | 82-92% |
EDTA plasma(n=5) | 81-101% | 79-101% | 84-104% | 79-102% |
heparin plasma(n=5) | 86-104% | 80-93% | 80-94% | 78-92% |
Stability of the ELISA Kit for Leptin (LEP)
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Assay procedure summary of the ELISA Kit for Leptin (LEP)
1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.
Test principle of the ELISA Kit for Leptin (LEP)
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Leptin (LEP). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Leptin (LEP). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Leptin (LEP), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Leptin (LEP) in the samples is then determined by comparing the O.D. of the samples to the standard curve.
GIVEAWAYS
INCREMENT SERVICES
Single-component Reagents of Assay Kit
Lysis Buffer Specific for ELISA / CLIA
Quality Control of Kit
ELISA Kit Customized Service
Disease Model Customized Service
Serums Customized Service
TGFB1 Activation Reagent
Real Time PCR Experimental Service
Streptavidin
Fast blue Protein Stain solution
Single-component Reagents of FLIA Kit
Streptavidin-Agarose Beads
Related products
Catalog No. | Organism species: Ovis aries; Ovine (Sheep) | Applications (RESEARCH USE ONLY!) |
RPA084Ov01 | Recombinant Leptin (LEP) | Positive Control; Immunogen; SDS-PAGE; WB. |
PAA084Ov01 | Polyclonal Antibody to Leptin (LEP) | WB; IHC; ICC; IP. |
SEA084Ov | ELISA Kit for Leptin (LEP) | Enzyme-linked immunosorbent assay for Antigen Detection. |
LMA084Ov | Multiplex Assay Kit for Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay) | FLIA Kit for Antigen Detection. |