ELISA Kit for Anti-Triiodothyronine Antibody (Anti-T3)

3,3',5-Triiodo-L-Thyronine; Liothyronine; Cytomel; Tertroxin

  • ELISA Kit for Anti-Triiodothyronine Antibody (Anti-T3) Packages (Simulation)
  • ELISA Kit for Anti-Triiodothyronine Antibody (Anti-T3) Packages (Simulation)
  • ELISA Kit for Anti-Triiodothyronine Antibody (Anti-T3) Results demonstration
  • AEA453Ge.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the ELISA Kit for Anti-Triiodothyronine Antibody (Anti-T3)

This assay has high sensitivity and excellent specificity for detection of Anti-Triiodothyronine Antibody (Anti-T3).
No significant cross-reactivity or interference between Anti-Triiodothyronine Antibody (Anti-T3) and analogues was observed.

Recovery of the ELISA Kit for Anti-Triiodothyronine Antibody (Anti-T3)

Matrices listed below were spiked with certain level of Anti-Triiodothyronine Antibody (Anti-T3) and the recovery rates were calculated by comparing the measured value to the expected amount of Anti-Triiodothyronine Antibody (Anti-T3) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 80-94 81
EDTA plasma(n=5) 93-102 98
heparin plasma(n=5) 80-96 90

Precision of the ELISA Kit for Anti-Triiodothyronine Antibody (Anti-T3)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Anti-Triiodothyronine Antibody (Anti-T3) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Anti-Triiodothyronine Antibody (Anti-T3) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the ELISA Kit for Anti-Triiodothyronine Antibody (Anti-T3)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Anti-Triiodothyronine Antibody (Anti-T3) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 85-92% 80-98% 88-99% 79-103%
EDTA plasma(n=5) 96-105% 88-102% 88-97% 78-105%
heparin plasma(n=5) 92-103% 94-105% 78-97% 95-102%

Stability of the ELISA Kit for Anti-Triiodothyronine Antibody (Anti-T3)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the ELISA Kit for Anti-Triiodothyronine Antibody (Anti-T3)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 5 times;
5. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
6. Add 50µL Stop Solution. Read at 450nm immediately.

Test principle of the ELISA Kit for Anti-Triiodothyronine Antibody (Anti-T3)

The microtiter plate provided in this kit has been pre-coated with an antigen. Standards or samples are then added to the appropriate microtiter plate wells with a Horseradish Peroxidase (HRP)-conjugated secondary antibody. After TMB substrate solution is added, those wells that contain Anti-Triiodothyronine Antibody (Anti-T3) will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Anti-Triiodothyronine Antibody (Anti-T3) in the samples is then determined by comparing the O.D. of the samples to the standard curve.

Related products

Catalog No. Organism species: Pan-species (General) Applications (RESEARCH USE ONLY!)
UCPA453Ge11 BSA Conjugated Triiodothyronine (T3) Immunogen; SDS-PAGE; WB.
UCPA453Ge21 OVA Conjugated Triiodothyronine (T3) Immunogen; SDS-PAGE; WB.
UPAA453Ge01 Polyclonal Antibody to Triiodothyronine (T3) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
ULAA453Ge71 Biotin-Linked Polyclonal Antibody to Triiodothyronine (T3) WB; IHC; ICC.
UMAA453Ge21 Monoclonal Antibody to Triiodothyronine (T3) WB; IHC; ICC; IP.
UMAA453Ge23 Monoclonal Antibody to Triiodothyronine (T3) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
UMAA453Ge22 Monoclonal Antibody to Triiodothyronine (T3) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
UMAA453Ge24 Monoclonal Antibody to Triiodothyronine (T3) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
UMAA453Ge25 Monoclonal Antibody to Triiodothyronine (T3) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
UMAA453Ge27 Monoclonal Antibody to Triiodothyronine (T3) WB; IHC; ICC; IP.
ULAA453Ge72 Biotin-Linked Monoclonal Antibody to Triiodothyronine (T3) WB; IHC; ICC.
UCEA453Ge ELISA Kit for Triiodothyronine (T3) Enzyme-linked immunosorbent assay for Antigen Detection.
UIEA453Ge Instant ELISA Kit for Triiodothyronine (T3) Enzyme-linked immunosorbent assay for Antigen Detection.
UMEA453Ge Mini Samples ELISA Kit for Triiodothyronine (T3) Enzyme-linked immunosorbent assay for Antigen Detection.
UAEA453Ge ELISA Kit for Anti-Triiodothyronine Antibody (Anti-T3) Enzyme-linked immunosorbent assay for Antibody Detection.
UCCA453Ge CLIA Kit for Triiodothyronine (T3) Chemiluminescent immunoassay for Antigen Detection.
ULMA453Ge Multiplex Assay Kit for Triiodothyronine (T3) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
UKSA453Ge11 ELISA Kit DIY Materials for Triiodothyronine (T3) Main materials for "Do It (ELISA Kit) Yourself".
GUESTBOOK