CLIA Kit for Flotillin 2 (FLOT2)

ECS1; ESA; ESA1; M17S1; Epidermal Surface Antigen 1; Membrane Component,Chromosome 17,Surface Marker 1; 35kD Protein Identified By Monoclonal Antibody ECS-1

  • CLIA Kit for Flotillin 2 (FLOT2) Packages (Simulation)
  • CLIA Kit for Flotillin 2 (FLOT2) Packages (Simulation)
  • CLIA Kit for Flotillin 2 (FLOT2) Results demonstration
  • SCC480Hu.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the CLIA Kit for Flotillin 2 (FLOT2)

This assay has high sensitivity and excellent specificity for detection of Flotillin 2 (FLOT2).
No significant cross-reactivity or interference between Flotillin 2 (FLOT2) and analogues was observed.

Precision of the CLIA Kit for Flotillin 2 (FLOT2)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Flotillin 2 (FLOT2) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Flotillin 2 (FLOT2) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Stability of the CLIA Kit for Flotillin 2 (FLOT2)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the CLIA Kit for Flotillin 2 (FLOT2)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

Test principle of the CLIA Kit for Flotillin 2 (FLOT2)

The microplate provided in this kit has been pre-coated with an antibody specific to Flotillin 2 (FLOT2). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Flotillin 2 (FLOT2). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Flotillin 2 (FLOT2) level in the sample or standard.;

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
URPC480Hu01 Recombinant Flotillin 2 (FLOT2) Positive Control; Immunogen; SDS-PAGE; WB.
UPAC480Hu01 Polyclonal Antibody to Flotillin 2 (FLOT2) WB
UMAC480Hu22 Monoclonal Antibody to Flotillin 2 (FLOT2) WB; IHC; ICC; IP.
USEC480Hu ELISA Kit for Flotillin 2 (FLOT2) Enzyme-linked immunosorbent assay for Antigen Detection.
USCC480Hu CLIA Kit for Flotillin 2 (FLOT2) Chemiluminescent immunoassay for Antigen Detection.
ULMC480Hu Multiplex Assay Kit for Flotillin 2 (FLOT2) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.