CLIA Kit for Fibulin 1 (FBLN1)

  • CLIA Kit for Fibulin 1 (FBLN1) Packages (Simulation)
  • CLIA Kit for Fibulin 1 (FBLN1) Packages (Simulation)
  • CLIA Kit for Fibulin 1 (FBLN1) Results demonstration
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the CLIA Kit for Fibulin 1 (FBLN1)

This assay has high sensitivity and excellent specificity for detection of Fibulin 1 (FBLN1).
No significant cross-reactivity or interference between Fibulin 1 (FBLN1) and analogues was observed.

Recovery of the CLIA Kit for Fibulin 1 (FBLN1)

Matrices listed below were spiked with certain level of recombinant Fibulin 1 (FBLN1) and the recovery rates were calculated by comparing the measured value to the expected amount of Fibulin 1 (FBLN1) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 86-95 91
EDTA plasma(n=5) 92-104 99
heparin plasma(n=5) 78-97 84

Precision of the CLIA Kit for Fibulin 1 (FBLN1)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Fibulin 1 (FBLN1) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Fibulin 1 (FBLN1) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the CLIA Kit for Fibulin 1 (FBLN1)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Fibulin 1 (FBLN1) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 94-102% 88-96% 81-104% 85-96%
EDTA plasma(n=5) 90-104% 96-103% 79-105% 78-103%
heparin plasma(n=5) 91-101% 95-105% 95-102% 96-104%

Stability of the CLIA Kit for Fibulin 1 (FBLN1)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the CLIA Kit for Fibulin 1 (FBLN1)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

Test principle of the CLIA Kit for Fibulin 1 (FBLN1)

The microplate provided in this kit has been pre-coated with an antibody specific to Fibulin 1 (FBLN1). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Fibulin 1 (FBLN1). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Fibulin 1 (FBLN1) level in the sample or standard.;

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
URPC472Hu02 Recombinant Fibulin 1 (FBLN1) Positive Control; Immunogen; SDS-PAGE; WB.
URPC472Hu03 Recombinant Fibulin 1 (FBLN1) Positive Control; Immunogen; SDS-PAGE; WB.
URPC472Hu01 Recombinant Fibulin 1 (FBLN1) Positive Control; Immunogen; SDS-PAGE; WB.
UPAC472Hu02 Polyclonal Antibody to Fibulin 1 (FBLN1) WB; IHC; ICC; IP.
UPAC472Hu01 Polyclonal Antibody to Fibulin 1 (FBLN1) WB; IHC; ICC; IP.
UPAC472Hu03 Polyclonal Antibody to Fibulin 1 (FBLN1) WB
UFAC472Hu01 Anti-Fibulin 1 (FBLN1) Polyclonal Antibody Flow cytometry.
UFAC472Hu03 Anti-Fibulin 1 (FBLN1) Polyclonal Antibody Flow cytometry.
ULAC472Hu71 Biotin-Linked Polyclonal Antibody to Fibulin 1 (FBLN1) WB; IHC; ICC.
UMAC472Hu21 Monoclonal Antibody to Fibulin 1 (FBLN1) WB; IHC; ICC; IP.
UMAC472Hu22 Monoclonal Antibody to Fibulin 1 (FBLN1) WB; IHC; ICC; IP.
UFAC472Hu02 Anti-Fibulin 1 (FBLN1) Monoclonal Antibody Flow cytometry.
ULAC472Hu72 Biotin-Linked Monoclonal Antibody to Fibulin 1 (FBLN1) WB; IHC; ICC.
USEC472Hu ELISA Kit for Fibulin 1 (FBLN1) Enzyme-linked immunosorbent assay for Antigen Detection.
USCC472Hu CLIA Kit for Fibulin 1 (FBLN1) Chemiluminescent immunoassay for Antigen Detection.
ULMC472Hu Multiplex Assay Kit for Fibulin 1 (FBLN1) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
UPSC472Hu01 Antibody Pair for Fibulin 1 (FBLN1) ELISA; CLIA; ELISPOT; Luminex; Immunochromatography and other Immunoassays.