Multiplex Assay Kit for Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay)

ARMET; ARP; Arginine Rich Mutated In Early Stage Tumors; Arginine-rich protein

(Note: Up to 8-plex in one testing reaction)

  • Multiplex Assay Kit for Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay) Packages (Simulation)
  • Multiplex Assay Kit for Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay) Packages (Simulation)
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the Multiplex Assay Kit for Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay)

This assay has high sensitivity and excellent specificity for detection of Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery of the Multiplex Assay Kit for Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay)

Matrices listed below were spiked with certain level of recombinant Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 88-101 95
EDTA plasma(n=5) 81-90 85
heparin plasma(n=5) 82-103 92

Precision of the Multiplex Assay Kit for Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the Multiplex Assay Kit for Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 92-99% 97-104% 82-96% 96-105%
EDTA plasma(n=5) 84-93% 82-96% 80-90% 96-104%
heparin plasma(n=5) 97-105% 82-92% 80-94% 94-103%

Stability of the Multiplex Assay Kit for Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the Multiplex Assay Kit for Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay)

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

Test principle of the Multiplex Assay Kit for Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay)

Analyte-specific antibodies are pre-coated onto color-coded microparticles. Microparticles, standards, and samples are pipetted into wells and the immobilized antibodies bind the analytes of interest. After washing away any unbound substances, a biotinylated antibody cocktail specific to the analytes of interest is added to each well. Following a wash to remove any unbound biotinylated antibody, Streptavidin-Phycoerythrin conjugate (Streptavidin-PE), which binds to the biotinylated detection antibodies, is added to each well. A final wash removes unbound Streptavidin-PE and the microparticles are resuspended in buffer and read using the Luminex or Bio-Plex analyzer.The MFI developed is proportional to the concentration of analytes of interest in the sample.

GIVEAWAYS

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
APC300Hu61 Active Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) Cell culture; Activity Assays.
EPC300Hu61 Eukaryotic Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) Positive Control; Immunogen; SDS-PAGE; WB.
APC300Hu01 Active Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) Cell culture; Activity Assays.
RPC300Hu01 Recombinant Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) Positive Control; Immunogen; SDS-PAGE; WB.
PAC300Hu01 Polyclonal Antibody to Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) WB; IHC; ICC; IP.
MAC300Hu25 Monoclonal Antibody to Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) WB
MAC300Hu21 Monoclonal Antibody to Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) WB; IHC; ICC; IP.
MAC300Hu23 Monoclonal Antibody to Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) WB; IHC; ICC; IP.
MAC300Hu24 Monoclonal Antibody to Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) WB; IHC; ICC; IP.
MAC300Hu27 Monoclonal Antibody to Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) WB; IHC; ICC; IP.
MAC300Hu22 Monoclonal Antibody to Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) WB,IHC
MEC300Hu Mini Samples ELISA Kit for Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) Enzyme-linked immunosorbent assay for Antigen Detection.
SEC300Hu ELISA Kit for Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) Enzyme-linked immunosorbent assay for Antigen Detection.
LMC300Hu Multiplex Assay Kit for Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSC300Hu01 ELISA Kit DIY Materials for Mesencephalic Astrocyte Derived Neurotrophic Factor (MANF) Main materials for "Do It(ELISA Kit) Yourself"
GUESTBOOK