Multiplex Assay Kit for Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay)

KIM-3; TIM3; TIMD3; Tim-3; T-cell membrane protein 3; T-cell immunoglobulin mucin receptor 3; T-cell immunoglobulin and mucin domain-containing protein 3

(Note: Up to 8-plex in one testing reaction)

  • Multiplex Assay Kit for Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay) Packages (Simulation)
  • Multiplex Assay Kit for Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay) Packages (Simulation)
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the Multiplex Assay Kit for Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay)

This assay has high sensitivity and excellent specificity for detection of Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery of the Multiplex Assay Kit for Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay)

Matrices listed below were spiked with certain level of recombinant Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 92-105 96
EDTA plasma(n=5) 79-102 97
heparin plasma(n=5) 96-104 101

Precision of the Multiplex Assay Kit for Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the Multiplex Assay Kit for Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 86-105% 93-102% 98-105% 81-93%
EDTA plasma(n=5) 98-105% 93-101% 86-93% 94-101%
heparin plasma(n=5) 80-97% 82-95% 87-102% 78-104%

Stability of the Multiplex Assay Kit for Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the Multiplex Assay Kit for Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay)

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

Test principle of the Multiplex Assay Kit for Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay)

Analyte-specific antibodies are pre-coated onto color-coded microparticles. Microparticles, standards, and samples are pipetted into wells and the immobilized antibodies bind the analytes of interest. After washing away any unbound substances, a biotinylated antibody cocktail specific to the analytes of interest is added to each well. Following a wash to remove any unbound biotinylated antibody, Streptavidin-Phycoerythrin conjugate (Streptavidin-PE), which binds to the biotinylated detection antibodies, is added to each well. A final wash removes unbound Streptavidin-PE and the microparticles are resuspended in buffer and read using the Luminex or Bio-Plex analyzer.The MFI developed is proportional to the concentration of analytes of interest in the sample.

GIVEAWAYS

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
EPH930Hu61 Eukaryotic Hepatitis A Virus Cellular Receptor 2 (HAVCR2) Positive Control; Immunogen; SDS-PAGE; WB.
APH930Hu61 Active Hepatitis A Virus Cellular Receptor 2 (HAVCR2) Cell culture; Activity Assays.
RPH930Hu01 Recombinant Hepatitis A Virus Cellular Receptor 2 (HAVCR2) Positive Control; Immunogen; SDS-PAGE; WB.
PAH930Hu01 Polyclonal Antibody to Hepatitis A Virus Cellular Receptor 2 (HAVCR2) WB,FCM
FAH930Hu01 Anti-Hepatitis A Virus Cellular Receptor 2 (HAVCR2) Polyclonal Antibody Flow cytometry.
FAH930Hu41 PE-Linked Anti-Hepatitis A Virus Cellular Receptor 2 (HAVCR2) Polyclonal Antibody Flow cytometry.
FAH930Hu51 APC-Linked Anti-Hepatitis A Virus Cellular Receptor 2 (HAVCR2) Polyclonal Antibody Flow cytometry.
FAH930Hu81 FITC-Linked Anti-Hepatitis A Virus Cellular Receptor 2 (HAVCR2) Polyclonal Antibody Flow cytometry.
MAH930Hu22 Monoclonal Antibody to Hepatitis A Virus Cellular Receptor 2 (HAVCR2) WB; IHC; ICC; IP.
MAH930Hu23 Monoclonal Antibody to Hepatitis A Virus Cellular Receptor 2 (HAVCR2) WB; IHC; ICC; IP.
MAH930Hu24 Monoclonal Antibody to Hepatitis A Virus Cellular Receptor 2 (HAVCR2) WB; IHC; ICC; IP.
MAH930Hu25 Monoclonal Antibody to Hepatitis A Virus Cellular Receptor 2 (HAVCR2) WB; IHC; ICC; IP.
MAH930Hu27 Monoclonal Antibody to Hepatitis A Virus Cellular Receptor 2 (HAVCR2) WB; IHC; ICC; IP.
MAH930Hu21 Monoclonal Antibody to Hepatitis A Virus Cellular Receptor 2 (HAVCR2) WB; IHC; ICC; IP.
LAH930Hu72 Biotin-Linked Monoclonal Antibody to Hepatitis A Virus Cellular Receptor 2 (HAVCR2) WB; IHC; ICC.
SEH930Hu ELISA Kit for Hepatitis A Virus Cellular Receptor 2 (HAVCR2) Enzyme-linked immunosorbent assay for Antigen Detection.
LMH930Hu Multiplex Assay Kit for Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
PSH930Hu01 Antibody Pair for Hepatitis A Virus Cellular Receptor 2 (HAVCR2) ELISA; CLIA; ELISPOT; Luminex; Immunochromatography and other Immunoassays.
KSH930Hu01 ELISA Kit DIY Materials for Hepatitis A Virus Cellular Receptor 2 (HAVCR2) Main materials for "Do It(ELISA Kit) Yourself"
GUESTBOOK