ELISA Kit for Tight Junction Protein 1 (TJP1)

ZO1; Zona Occludens 1; Zona occludens protein 1; Zonula occludens protein 1

  • ELISA Kit for Tight Junction Protein 1 (TJP1) Packages (Simulation)
  • ELISA Kit for Tight Junction Protein 1 (TJP1) Packages (Simulation)
  • ELISA Kit for Tight Junction Protein 1 (TJP1) Results demonstration
  • SEC262Ra.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the ELISA Kit for Tight Junction Protein 1 (TJP1)

This assay has high sensitivity and excellent specificity for detection of Tight Junction Protein 1 (TJP1).
No significant cross-reactivity or interference between Tight Junction Protein 1 (TJP1) and analogues was observed.

Precision of the ELISA Kit for Tight Junction Protein 1 (TJP1)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Tight Junction Protein 1 (TJP1) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Tight Junction Protein 1 (TJP1) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Stability of the ELISA Kit for Tight Junction Protein 1 (TJP1)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the ELISA Kit for Tight Junction Protein 1 (TJP1)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

Test principle of the ELISA Kit for Tight Junction Protein 1 (TJP1)

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Tight Junction Protein 1 (TJP1). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Tight Junction Protein 1 (TJP1). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Tight Junction Protein 1 (TJP1), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Tight Junction Protein 1 (TJP1) in the samples is then determined by comparing the O.D. of the samples to the standard curve.

Related products

Catalog No. Organism species: Rattus norvegicus (Rat) Applications (RESEARCH USE ONLY!)
RPC262Ra01 Recombinant Tight Junction Protein 1 (TJP1) Positive Control; Immunogen; SDS-PAGE; WB.
PAC262Ra01 Polyclonal Antibody to Tight Junction Protein 1 (TJP1) WB; IHC; ICC; IP.
MAC262Ra21 Monoclonal Antibody to Tight Junction Protein 1 (TJP1) WB; IHC; ICC; IP.
SEC262Ra ELISA Kit for Tight Junction Protein 1 (TJP1) Enzyme-linked immunosorbent assay for Antigen Detection.
LMC262Ra Multiplex Assay Kit for Tight Junction Protein 1 (TJP1) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.