CLIA Kit for Alpha-1-Antitrypsin (a1AT)

SERPINA1; SPAAT; A1-AT; PI; Serpin Peptidase Inhibitor,Clade A(Alpha-1 Antiproteinase/AntiTrypsin)Member 1; Alpha-1 protease inhibitor; Short peptide from AAT; Serpin A1

  • CLIA Kit for Alpha-1-Antitrypsin (a1AT) Packages (Simulation)
  • CLIA Kit for Alpha-1-Antitrypsin (a1AT) Packages (Simulation)
  • CLIA Kit for Alpha-1-Antitrypsin (a1AT) Results demonstration
  • SCB697Hu.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the CLIA Kit for Alpha-1-Antitrypsin (a1AT)

This assay has high sensitivity and excellent specificity for detection of Alpha-1-Antitrypsin (a1AT).
No significant cross-reactivity or interference between Alpha-1-Antitrypsin (a1AT) and analogues was observed.

Recovery of the CLIA Kit for Alpha-1-Antitrypsin (a1AT)

Matrices listed below were spiked with certain level of recombinant Alpha-1-Antitrypsin (a1AT) and the recovery rates were calculated by comparing the measured value to the expected amount of Alpha-1-Antitrypsin (a1AT) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 78-91 80
EDTA plasma(n=5) 81-93 88
heparin plasma(n=5) 80-88 84

Precision of the CLIA Kit for Alpha-1-Antitrypsin (a1AT)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Alpha-1-Antitrypsin (a1AT) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Alpha-1-Antitrypsin (a1AT) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the CLIA Kit for Alpha-1-Antitrypsin (a1AT)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Alpha-1-Antitrypsin (a1AT) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 86-102% 80-95% 88-99% 81-101%
EDTA plasma(n=5) 82-105% 86-104% 83-101% 79-91%
heparin plasma(n=5) 97-105% 88-95% 78-90% 79-102%

Stability of the CLIA Kit for Alpha-1-Antitrypsin (a1AT)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the CLIA Kit for Alpha-1-Antitrypsin (a1AT)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

Test principle of the CLIA Kit for Alpha-1-Antitrypsin (a1AT)

The microplate provided in this kit has been pre-coated with an antibody specific to Alpha-1-Antitrypsin (a1AT). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Alpha-1-Antitrypsin (a1AT). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Alpha-1-Antitrypsin (a1AT) level in the sample or standard.;

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
RPB697Hu01 Recombinant Alpha-1-Antitrypsin (a1AT) Positive Control; Immunogen; SDS-PAGE; WB.
PAB697Hu01 Polyclonal Antibody to Alpha-1-Antitrypsin (a1AT) WB; IHC; ICC; IP.
LAB697Hu71 Biotin-Linked Polyclonal Antibody to Alpha-1-Antitrypsin (a1AT) WB; IHC; ICC.
MAB697Hu24 Monoclonal Antibody to Alpha-1-Antitrypsin (a1AT) WB; IHC; ICC; IP.
MAB697Hu25 Monoclonal Antibody to Alpha-1-Antitrypsin (a1AT) WB; IHC; ICC; IP.
MAB697Hu23 Monoclonal Antibody to Alpha-1-Antitrypsin (a1AT) WB; IHC; ICC; IP.
MAB697Hu29 Monoclonal Antibody to Alpha-1-Antitrypsin (a1AT) WB; IHC; ICC; IP.
MAB697Hu27 Monoclonal Antibody to Alpha-1-Antitrypsin (a1AT) WB; IHC; ICC; IP.
MAB697Hu26 Monoclonal Antibody to Alpha-1-Antitrypsin (a1AT) WB; IHC; ICC; IP.
MAB697Hu21 Monoclonal Antibody to Alpha-1-Antitrypsin (a1AT) WB; IHC; ICC/IF.
MAB697Hu22 Monoclonal Antibody to Alpha-1-Antitrypsin (a1AT) WB; IHC; ICC; IP.
FAB697Hu02 Anti-Alpha-1-Antitrypsin (a1AT) Monoclonal Antibody Flow cytometry.
MEB697Hu Mini Samples ELISA Kit for Alpha-1-Antitrypsin (a1AT) Enzyme-linked immunosorbent assay for Antigen Detection.
SEB697Hu ELISA Kit for Alpha-1-Antitrypsin (a1AT) Enzyme-linked immunosorbent assay for Antigen Detection.
SCB697Hu CLIA Kit for Alpha-1-Antitrypsin (a1AT) Chemiluminescent immunoassay for Antigen Detection.
LMB697Hu Multiplex Assay Kit for Alpha-1-Antitrypsin (a1AT) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSB697Hu01 ELISA Kit DIY Materials for Alpha-1-Antitrypsin (a1AT) Main materials for "Do It (ELISA Kit) Yourself".
GUESTBOOK