CLIA Kit for Activin A (ACVA)

ACV-A; Activin Beta A Beta A Homodimer

  • CLIA Kit for Activin A (ACVA) Packages (Simulation)
  • CLIA Kit for Activin A (ACVA) Packages (Simulation)
  • CLIA Kit for Activin A (ACVA) Results demonstration
  • SCA001Hu.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the CLIA Kit for Activin A (ACVA)

This assay has high sensitivity and excellent specificity for detection of Activin A (ACVA).
No significant cross-reactivity or interference between Activin A (ACVA) and analogues was observed.

Recovery of the CLIA Kit for Activin A (ACVA)

Matrices listed below were spiked with certain level of recombinant Activin A (ACVA) and the recovery rates were calculated by comparing the measured value to the expected amount of Activin A (ACVA) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 86-96 92
EDTA plasma(n=5) 97-105 101
heparin plasma(n=5) 97-105 102

Precision of the CLIA Kit for Activin A (ACVA)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Activin A (ACVA) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Activin A (ACVA) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the CLIA Kit for Activin A (ACVA)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Activin A (ACVA) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 98-105% 92-103% 87-101% 96-105%
EDTA plasma(n=5) 80-93% 88-95% 78-92% 99-105%
heparin plasma(n=5) 89-102% 82-104% 92-99% 83-97%

Stability of the CLIA Kit for Activin A (ACVA)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the CLIA Kit for Activin A (ACVA)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

Test principle of the CLIA Kit for Activin A (ACVA)

The microplate provided in this kit has been pre-coated with an antibody specific to Activin A (ACVA). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Activin A (ACVA). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Activin A (ACVA) level in the sample or standard.;

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
RPA001Hu01 Recombinant Activin A (ACVA) Positive Control; Immunogen; SDS-PAGE; WB.
APA001Hu01 Active Activin A (ACVA) Cell culture; Activity Assays.
PAA001Hu01 Polyclonal Antibody to Activin A (ACVA) WB; ICC; IHC-P; IHC-F; ELISA.
SEA001Hu ELISA Kit for Activin A (ACVA) Enzyme-linked immunosorbent assay for Antigen Detection.
CEA001Hu ELISA Kit for Activin A (ACVA) Enzyme-linked immunosorbent assay for Antigen Detection.
SCA001Hu CLIA Kit for Activin A (ACVA) Chemiluminescent immunoassay for Antigen Detection.
LMA001Hu Multiplex Assay Kit for Activin A (ACVA) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.