Wide-range ELISA Kit for Interleukin 4 (IL4)

BSF1; BCGF1 ;B Cell Stimulatory Factor 1; Lymphocyte stimulatory factor 1; B Cell Growth Factor; Binetrakin; Pitrakinra

  • Wide-range ELISA Kit for Interleukin 4 (IL4) Packages (Simulation)
  • Wide-range ELISA Kit for Interleukin 4 (IL4) Packages (Simulation)
  • Wide-range ELISA Kit for Interleukin 4 (IL4) Results demonstration
  • WEA077Mu.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the Wide-range ELISA Kit for Interleukin 4 (IL4)

This assay has high sensitivity and excellent specificity for detection of Wide-range Interleukin 4 (IL4).
No significant cross-reactivity or interference between Wide-range Interleukin 4 (IL4) and analogues was observed.

Recovery of the Wide-range ELISA Kit for Interleukin 4 (IL4)

Matrices listed below were spiked with certain level of recombinant Wide-range Interleukin 4 (IL4) and the recovery rates were calculated by comparing the measured value to the expected amount of Wide-range Interleukin 4 (IL4) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 95-105 99
EDTA plasma(n=5) 78-101 82
heparin plasma(n=5) 87-101 91

Precision of the Wide-range ELISA Kit for Interleukin 4 (IL4)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Wide-range Interleukin 4 (IL4) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Wide-range Interleukin 4 (IL4) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the Wide-range ELISA Kit for Interleukin 4 (IL4)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Wide-range Interleukin 4 (IL4) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 92-105% 80-101% 99-105% 95-102%
EDTA plasma(n=5) 93-101% 78-94% 79-95% 81-104%
heparin plasma(n=5) 78-99% 91-102% 80-90% 94-105%

Stability of the Wide-range ELISA Kit for Interleukin 4 (IL4)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the Wide-range ELISA Kit for Interleukin 4 (IL4)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

Test principle of the Wide-range ELISA Kit for Interleukin 4 (IL4)

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Wide-range Interleukin 4 (IL4). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Wide-range Interleukin 4 (IL4). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Wide-range Interleukin 4 (IL4), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Wide-range Interleukin 4 (IL4) in the samples is then determined by comparing the O.D. of the samples to the standard curve.

Related products

Catalog No. Organism species: Mus musculus (Mouse) Applications (RESEARCH USE ONLY!)
RPA077Mu02 Recombinant Interleukin 4 (IL4) Positive Control; Immunogen; SDS-PAGE; WB.
EPA077Mu61 Eukaryotic Interleukin 4 (IL4) Positive Control; Immunogen; SDS-PAGE; WB.
RPA077Mu01 Recombinant Interleukin 4 (IL4) Positive Control; Immunogen; SDS-PAGE; WB.
APA077Mu01 Active Interleukin 4 (IL4) Cell culture; Activity Assays.
APA077Mu61 Active Interleukin 4 (IL4) Cell culture; Activity Assays.
PAA077Mu06 Polyclonal Antibody to Interleukin 4 (IL4) WB; IHC; ICC; IP.
PAA077Mu02 Polyclonal Antibody to Interleukin 4 (IL4) WB; IHC; ICC; IP.
PAA077Mu01 Polyclonal Antibody to Interleukin 4 (IL4) WB; IHC; ICC; IP.
WEA077Mu Wide-range ELISA Kit for Interleukin 4 (IL4) Enzyme-linked immunosorbent assay for Antigen Detection.
MEA077Mu Mini Samples ELISA Kit for Interleukin 4 (IL4) Enzyme-linked immunosorbent assay for Antigen Detection.
SEA077Mu ELISA Kit for Interleukin 4 (IL4) Enzyme-linked immunosorbent assay for Antigen Detection.
SCA077Mu CLIA Kit for Interleukin 4 (IL4) Chemiluminescent immunoassay for Antigen Detection.
LMA077Mu Multiplex Assay Kit for Interleukin 4 (IL4) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA077Mu01 ELISA Kit DIY Materials for Interleukin 4 (IL4) Main materials for "Do It (ELISA Kit) Yourself".
GUESTBOOK