Multiplex Assay Kit for Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay)

OB; OBS; Obesity Homolog; Obesity Factor; Obese Protein

(Note: Up to 8-plex in one testing reaction)

  • Multiplex Assay Kit for Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay) Packages (Simulation)
  • Multiplex Assay Kit for Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay) Packages (Simulation)
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the Multiplex Assay Kit for Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay)

This assay has high sensitivity and excellent specificity for detection of Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery of the Multiplex Assay Kit for Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay)

Matrices listed below were spiked with certain level of recombinant Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 79-96 79
EDTA plasma(n=5) 84-94 90
heparin plasma(n=5) 79-93 88

Precision of the Multiplex Assay Kit for Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the Multiplex Assay Kit for Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 82-102% 79-91% 95-103% 81-89%
EDTA plasma(n=5) 95-102% 81-93% 83-92% 80-103%
heparin plasma(n=5) 82-99% 98-105% 90-101% 83-97%

Stability of the Multiplex Assay Kit for Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the Multiplex Assay Kit for Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay)

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

Test principle of the Multiplex Assay Kit for Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay)

Analyte-specific antibodies are pre-coated onto color-coded microparticles. Microparticles, standards, and samples are pipetted into wells and the immobilized antibodies bind the analytes of interest. After washing away any unbound substances, a biotinylated antibody cocktail specific to the analytes of interest is added to each well. Following a wash to remove any unbound biotinylated antibody, Streptavidin-Phycoerythrin conjugate (Streptavidin-PE), which binds to the biotinylated detection antibodies, is added to each well. A final wash removes unbound Streptavidin-PE and the microparticles are resuspended in buffer and read using the Luminex or Bio-Plex analyzer.The MFI developed is proportional to the concentration of analytes of interest in the sample.

GIVEAWAYS

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
APA084Hu01 Active Leptin (LEP) Cell culture; Activity Assays.
RPA084Hu01 Recombinant Leptin (LEP) Positive Control; Immunogen; SDS-PAGE; WB.
PAA084Hu01 Polyclonal Antibody to Leptin (LEP) WB; IHC; ICC; IP.
LAA084Hu71 Biotin-Linked Polyclonal Antibody to Leptin (LEP) WB; IHC; ICC.
LAA084Hu81 FITC-Linked Polyclonal Antibody to Leptin (LEP) WB; IHC; ICC; IF.
MAA084Hu22 Monoclonal Antibody to Leptin (LEP) WB; IHC; ICC; IP.
MAA084Hu21 Monoclonal Antibody to Leptin (LEP) WB
MAA084Hu23 Monoclonal Antibody to Leptin (LEP) WB; IHC; ICC; IP.
MAA084Hu24 Monoclonal Antibody to Leptin (LEP) WB; IHC; ICC; IP.
MAA084Hu25 Monoclonal Antibody to Leptin (LEP) WB; IHC; ICC; IP.
HEA084Hu High Sensitive ELISA Kit for Leptin (LEP) Enzyme-linked immunosorbent assay for Antigen Detection.
SEA084Hu ELISA Kit for Leptin (LEP) Enzyme-linked immunosorbent assay for Antigen Detection.
MEA084Hu Mini Samples ELISA Kit for Leptin (LEP) Enzyme-linked immunosorbent assay for Antigen Detection.
SCA084Hu CLIA Kit for Leptin (LEP) Chemiluminescent immunoassay for Antigen Detection.
LMA084Hu Multiplex Assay Kit for Leptin (LEP) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA084Hu01 ELISA Kit DIY Materials for Leptin (LEP) Main materials for "Do It (ELISA Kit) Yourself".
GUESTBOOK