Multiplex Assay Kit for Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay)

(Note: Up to 8-plex in one testing reaction)

  • Multiplex Assay Kit for Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay) Packages (Simulation)
  • Multiplex Assay Kit for Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay) Packages (Simulation)
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the Multiplex Assay Kit for Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay)

This assay has high sensitivity and excellent specificity for detection of Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery of the Multiplex Assay Kit for Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay)

Matrices listed below were spiked with certain level of recombinant Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 90-99 95
EDTA plasma(n=5) 92-104 101
heparin plasma(n=5) 88-105 94
sodium citrate plasma(n=5) 85-95 91

Precision of the Multiplex Assay Kit for Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the Multiplex Assay Kit for Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 79-103% 79-90% 97-104% 79-94%
EDTA plasma(n=5) 80-103% 85-96% 94-103% 82-92%
heparin plasma(n=5) 85-99% 83-93% 79-101% 92-101%
sodium citrate plasma(n=5) 79-101% 88-96% 83-98% 86-102%

Stability of the Multiplex Assay Kit for Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the Multiplex Assay Kit for Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay)

1. Preparation of standards, reagents and samples before the experiment;
2. Add 50μL standard or sample to each well,
    add 10μL magnetic beads,and 50μL Detection Reagent A,incubate 60min at 37°C on shaker;
3. Wash plate on magnetic frame for three times;
4. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
5. Wash plate on magnetic frame for three times;
6. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

Test principle of the Multiplex Assay Kit for Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay)

Analyte-specific antibodies are pre-coated onto color-coded microparticles. Microparticles, standards,Labeled antigen and samples are pipetted into wells and the immobilized antibodies bind the analytes of interest.A competitive inhibition reaction is launched between biotin labeled analytes of interest and unlabeled analytes of interest (Standards or samples) with the pre-coated antibody specific to analytes of interest. Following a wash to remove any unbound substances, Streptavidin-Phycoerythrin conjugate (Streptavidin-PE) is added to each well. A final wash removes unbound Streptavidin-PE and the microparticles are resuspended in buffer and read using the Luminex or Bio-Plex analyzer. The MFI developed is reverse proportional to the concentration of analytes of interest in the sample.

GIVEAWAYS

INCREMENT SERVICES

Related products

Catalog No. Organism species: Oryctolagus cuniculus (Rabbit) Applications (RESEARCH USE ONLY!)
NPA544Rb01 Native Immunoglobulin G (IgG) Positive Control; Immunogen; SDS-PAGE; WB.
EPA544Rb61 Eukaryotic Immunoglobulin G (IgG) Positive Control; Immunogen; SDS-PAGE; WB.
SAA544Rb10 None-Linked Caprine Anti-Rabbit IgG Polyclonal Antibody WB; IHC; ICC; IF.
PAA544Rb51 Polyclonal Antibody to Immunoglobulin G (IgG) WB; IHC; ICC; IP.
SAA544Rb14 PE-Linked Caprine Anti-Rabbit IgG Polyclonal Antibody WB; IHC; ICC; IF.
SAA544Rb50 None-Linked Guinea pig Anti-Rabbit IgG Polyclonal Antibody WB; IP; IHC-P; IHC-F; ELISA.
SAA544Rb16 Cy3-Linked Caprine Anti-Rabbit IgG Polyclonal Antibody WB; IHC; ICC; IF.
SAA544Rb56 Cy3-Linked Guinea pig Anti-Rabbit IgG Polyclonal Antibody WB; IHC; ICC; IF.
SAA544Rb58 FITC-Linked Guinea pig Anti-Rabbit IgG Polyclonal Antibody WB; IHC; ICC; IF.
SAA544Rb15 APC-Linked Caprine Anti-Rabbit IgG Polyclonal Antibody WB; IHC; ICC; IF.
SAA544Rb17 Biotin-Linked Caprine Anti-Rabbit IgG Polyclonal Antibody WB; IHC; ICC; IF.
SAA544Rb18 FITC-Linked Caprine Anti-Rabbit IgG Polyclonal Antibody WB; IHC; ICC; IF.
SAA544Rb19 HRP-Linked Caprine Anti-Rabbit IgG Polyclonal Antibody WB; IHC; ICC.
SAA544Rb54 PE-Linked Guinea pig Anti-Rabbit IgG Polyclonal Antibody WB; IHC; ICC; IF.
SAA544Rb57 Biotin-Linked Guinea pig Anti-Rabbit IgG Polyclonal Antibody WB; IHC; ICC.
SAA544Rb55 APC-Linked Guinea pig Anti-Rabbit IgG Polyclonal Antibody WB; IHC; ICC; IF.
SAA544Rb59 HRP-Linked Guinea pig Anti-Rabbit IgG Polyclonal Antibody WB; IHC; ICC.
MAA544Rb27 Monoclonal Antibody to Immunoglobulin G (IgG) WB; IHC; ICC; IP.
MAA544Rb29 Monoclonal Antibody to Immunoglobulin G (IgG) WB; IHC; ICC; IP.
MAA544Rb21 Monoclonal Antibody to Immunoglobulin G (IgG) WB; IHC; ICC; IP.
MAA544Rb22 Monoclonal Antibody to Immunoglobulin G (IgG) WB; IHC; ICC; IP.
MAA544Rb23 Monoclonal Antibody to Immunoglobulin G (IgG) WB; IHC; ICC; IP.
MAA544Rb24 Monoclonal Antibody to Immunoglobulin G (IgG) WB; IHC; ICC; IP.
MAA544Rb25 Monoclonal Antibody to Immunoglobulin G (IgG) WB; IHC; ICC; IP.
CEA544Rb ELISA Kit for Immunoglobulin G (IgG) Enzyme-linked immunosorbent assay for Antigen Detection.
CCA544Rb CLIA Kit for Immunoglobulin G (IgG) Chemiluminescent immunoassay for Antigen Detection.
LMA544Rb Multiplex Assay Kit for Immunoglobulin G (IgG) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA544Rb11 ELISA Kit DIY Materials for Immunoglobulin G (IgG) Main materials for "Do It (ELISA Kit) Yourself".
GUESTBOOK