Multiplex Assay Kit for Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay)

ENO2; Enolase 2; Gamma Enolase; 2-phospho-D-glycerate hydro-lyase; Neural enolase

(Note: Up to 8-plex in one testing reaction)

  • Multiplex Assay Kit for Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay) Packages (Simulation)
  • Multiplex Assay Kit for Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay) Packages (Simulation)
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the Multiplex Assay Kit for Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay)

This assay has high sensitivity and excellent specificity for detection of Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay).
No significant cross-reactivity or interference between Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay) and analogues was observed.

Recovery of the Multiplex Assay Kit for Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay)

Matrices listed below were spiked with certain level of recombinant Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay) and the recovery rates were calculated by comparing the measured value to the expected amount of Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 84-103 92
EDTA plasma(n=5) 80-92 87
heparin plasma(n=5) 90-97 93

Precision of the Multiplex Assay Kit for Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the Multiplex Assay Kit for Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 87-101% 83-103% 94-101% 84-93%
EDTA plasma(n=5) 97-105% 85-104% 98-105% 96-103%
heparin plasma(n=5) 90-101% 98-105% 84-103% 96-103%

Stability of the Multiplex Assay Kit for Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the Multiplex Assay Kit for Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay)

1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
    add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.

Test principle of the Multiplex Assay Kit for Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay)

Analyte-specific antibodies are pre-coated onto color-coded microparticles. Microparticles, standards, and samples are pipetted into wells and the immobilized antibodies bind the analytes of interest. After washing away any unbound substances, a biotinylated antibody cocktail specific to the analytes of interest is added to each well. Following a wash to remove any unbound biotinylated antibody, Streptavidin-Phycoerythrin conjugate (Streptavidin-PE), which binds to the biotinylated detection antibodies, is added to each well. A final wash removes unbound Streptavidin-PE and the microparticles are resuspended in buffer and read using the Luminex or Bio-Plex analyzer.The MFI developed is proportional to the concentration of analytes of interest in the sample.

GIVEAWAYS

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
RPA537Hu02 Recombinant Enolase, Neuron Specific (NSE) Positive Control; Immunogen; SDS-PAGE; WB.
RPA537Hu01 Recombinant Enolase, Neuron Specific (NSE) Positive Control; Immunogen; SDS-PAGE; WB.
PAA537Hu01 Polyclonal Antibody to Enolase, Neuron Specific (NSE) WB,IHC
PAA537Hu02 Polyclonal Antibody to Enolase, Neuron Specific (NSE) WB; IHC; ICC; IP.
LAA537Hu71 Biotin-Linked Polyclonal Antibody to Enolase, Neuron Specific (NSE) WB; IHC; ICC.
LAA537Hu81 FITC-Linked Polyclonal Antibody to Enolase, Neuron Specific (NSE) WB; IHC; ICC; IF.
MAA537Hu27 Monoclonal Antibody to Enolase, Neuron Specific (NSE) WB; IHC; ICC; IP.
MAA537Hu29 Monoclonal Antibody to Enolase, Neuron Specific (NSE) WB; IHC; ICC; IP.
MAA537Hu25 Monoclonal Antibody to Enolase, Neuron Specific (NSE) WB; IHC; ICC; IP.
MAA537Hu21 Monoclonal Antibody to Enolase, Neuron Specific (NSE) WB; IHC; ICC; IP.
MAA537Hu22 Monoclonal Antibody to Enolase, Neuron Specific (NSE) WB; IHC; ICC; IP.
MAA537Hu23 Monoclonal Antibody to Enolase, Neuron Specific (NSE) WB; IHC; ICC; IP.
MAA537Hu26 Monoclonal Antibody to Enolase, Neuron Specific (NSE) WB; IHC; ICC; IP.
MAA537Hu28 Monoclonal Antibody to Enolase, Neuron Specific (NSE) WB; IHC; ICC; IP.
MAA537Hu24 Monoclonal Antibody to Enolase, Neuron Specific (NSE) WB; IHC; ICC; IP.
SEA537Hu ELISA Kit for Enolase, Neuron Specific (NSE) Enzyme-linked immunosorbent assay for Antigen Detection.
LMA537Hu Multiplex Assay Kit for Enolase, Neuron Specific (NSE) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA537Hu01 ELISA Kit DIY Materials for Enolase, Neuron Specific (NSE) Main materials for "Do It (ELISA Kit) Yourself".
CGA537Hu01 cDNA of Enolase, Neuron Specific (NSE) PCR; Protein Expression; etc.
GUESTBOOK