High Sensitive ELISA Kit for Interleukin 17 (IL17)

IL17A; CTLA8; IL-17A; Cytotoxic T-Lymphocyte-Associated Protein 8

  • High Sensitive ELISA Kit for Interleukin 17 (IL17) Packages (Simulation)
  • High Sensitive ELISA Kit for Interleukin 17 (IL17) Packages (Simulation)
  • High Sensitive ELISA Kit for Interleukin 17 (IL17) Results demonstration
  • HEA063Hu.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the High Sensitive ELISA Kit for Interleukin 17 (IL17)

This assay has high sensitivity and excellent specificity for detection of High Sensitive Interleukin 17 (IL17).
No significant cross-reactivity or interference between High Sensitive Interleukin 17 (IL17) and analogues was observed.

Recovery of the High Sensitive ELISA Kit for Interleukin 17 (IL17)

Matrices listed below were spiked with certain level of recombinant High Sensitive Interleukin 17 (IL17) and the recovery rates were calculated by comparing the measured value to the expected amount of High Sensitive Interleukin 17 (IL17) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 88-101 96
EDTA plasma(n=5) 82-96 86
heparin plasma(n=5) 84-92 88

Precision of the High Sensitive ELISA Kit for Interleukin 17 (IL17)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level High Sensitive Interleukin 17 (IL17) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level High Sensitive Interleukin 17 (IL17) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the High Sensitive ELISA Kit for Interleukin 17 (IL17)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of High Sensitive Interleukin 17 (IL17) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 80-92% 91-101% 80-96% 84-96%
EDTA plasma(n=5) 83-91% 92-103% 81-105% 91-105%
heparin plasma(n=5) 93-103% 89-97% 96-103% 92-102%

Stability of the High Sensitive ELISA Kit for Interleukin 17 (IL17)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the High Sensitive ELISA Kit for Interleukin 17 (IL17)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

Test principle of the High Sensitive ELISA Kit for Interleukin 17 (IL17)

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to High Sensitive Interleukin 17 (IL17). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to High Sensitive Interleukin 17 (IL17). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain High Sensitive Interleukin 17 (IL17), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of High Sensitive Interleukin 17 (IL17) in the samples is then determined by comparing the O.D. of the samples to the standard curve.

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
EPA063Hu61 Eukaryotic Interleukin 17 (IL17) Positive Control; Immunogen; SDS-PAGE; WB.
APA063Hu01 Active Interleukin 17 (IL17) Cell culture; Activity Assays.
RPA063Hu01 Recombinant Interleukin 17 (IL17) Positive Control; Immunogen; SDS-PAGE; WB.
PAA063Hu06 Polyclonal Antibody to Interleukin 17 (IL17) WB; IHC; ICC; IP.
PAA063Hu01 Polyclonal Antibody to Interleukin 17 (IL17) WB; IHC; ICC; IP.
RAA063Hu21 Recombinant Antibody to Interleukin 17 (IL17) WB; IF; ICC; IHC; IP; FCM.
MAA063Hu21 Monoclonal Antibody to Interleukin 17 (IL17) WB; IHC; ICC; IP.
MAA063Hu22 Monoclonal Antibody to Interleukin 17 (IL17) WB; IHC; ICC; IP.
MAA063Hu23 Monoclonal Antibody to Interleukin 17 (IL17) IHC
MAA063Hu24 Monoclonal Antibody to Interleukin 17 (IL17) WB; IHC; ICC; IP.
MAA063Hu25 Monoclonal Antibody to Interleukin 17 (IL17) WB; IHC; ICC; IP.
FAA063Hu12 Labelled Anti-Interleukin 17 (IL17) Monoclonal Antibody FCM
HEA063Hu High Sensitive ELISA Kit for Interleukin 17 (IL17) Enzyme-linked immunosorbent assay for Antigen Detection.
MEA063Hu Mini Samples ELISA Kit for Interleukin 17 (IL17) Enzyme-linked immunosorbent assay for Antigen Detection.
SEA063Hu ELISA Kit for Interleukin 17 (IL17) Enzyme-linked immunosorbent assay for Antigen Detection.
SCA063Hu CLIA Kit for Interleukin 17 (IL17) Chemiluminescent immunoassay for Antigen Detection.
LMA063Hu Multiplex Assay Kit for Interleukin 17 (IL17) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA063Hu01 ELISA Kit DIY Materials for Interleukin 17 (IL17) Main materials for "Do It(ELISA Kit) Yourself"