Specificity of the ELISA Kit for Vitamin B2 (VB2)

This assay has high sensitivity and excellent specificity for detection of Vitamin B2 (VB2).
No significant cross-reactivity or interference between Vitamin B2 (VB2) and analogues was observed.

Recovery of the ELISA Kit for Vitamin B2 (VB2)

Matrices listed below were spiked with certain level of Vitamin B2 (VB2) and the recovery rates were calculated by comparing the measured value to the expected amount of Vitamin B2 (VB2) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 79-101 87
EDTA plasma(n=5) 84-95 92
heparin plasma(n=5) 84-104 99

Precision of the ELISA Kit for Vitamin B2 (VB2)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Vitamin B2 (VB2) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Vitamin B2 (VB2) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the ELISA Kit for Vitamin B2 (VB2)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Vitamin B2 (VB2) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 80-104% 92-104% 80-95% 93-103%
EDTA plasma(n=5) 82-94% 83-95% 99-105% 84-98%
heparin plasma(n=5) 80-103% 82-92% 93-105% 78-96%

Stability of the ELISA Kit for Vitamin B2 (VB2)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the ELISA Kit for Vitamin B2 (VB2)

1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
    And then add 50µL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
7. Add 50µL Stop Solution. Read at 450 nm immediately.

Test principle of the ELISA Kit for Vitamin B2 (VB2)

This assay employs the competitive inhibition enzyme immunoassay technique. A monoclonal antibody specific to Vitamin B2 (VB2) has been pre-coated onto a microplate. A competitive inhibition reaction is launched between biotin labeled Vitamin B2 (VB2) and unlabeled Vitamin B2 (VB2) (Standards or samples) with the pre-coated antibody specific to Vitamin B2 (VB2). After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of Vitamin B2 (VB2) in the sample. After addition of the substrate solution, the intensity of color developed is reverse proportional to the concentration of Vitamin B2 (VB2) in the sample.

Related products

Catalog No. Organism species: Pan-species (General) Applications (RESEARCH USE ONLY!)
CPD054Ge11 BSA Conjugated Vitamin B2 (VB2) Immunogen; SDS-PAGE; WB.
CPD054Ge21 OVA Conjugated Vitamin B2 (VB2) Immunogen; SDS-PAGE; WB.
PAD054Ge01 Polyclonal Antibody to Vitamin B2 (VB2) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
LAD054Ge71 Biotin-Linked Polyclonal Antibody to Vitamin B2 (VB2) WB; IHC; ICC.
MAD054Ge23 Monoclonal Antibody to Vitamin B2 (VB2) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
MAD054Ge25 Monoclonal Antibody to Vitamin B2 (VB2) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
MAD054Ge26 Monoclonal Antibody to Vitamin B2 (VB2) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
MAD054Ge21 Monoclonal Antibody to Vitamin B2 (VB2) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
MAD054Ge22 Monoclonal Antibody to Vitamin B2 (VB2) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
MAD054Ge24 Monoclonal Antibody to Vitamin B2 (VB2) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
CED054Ge ELISA Kit for Vitamin B2 (VB2) Enzyme-linked immunosorbent assay for Antigen Detection.
LMD054Ge Multiplex Assay Kit for Vitamin B2 (VB2) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSD054Ge11 ELISA Kit DIY Materials for Vitamin B2 (VB2) Main materials for "Do It(ELISA Kit) Yourself"
GUESTBOOK