ELISA Kit for Superoxide Dismutase 1 (SOD1)

ALS; ALS1; IPOA; Superoxide Dismutase 1, Soluble; Amyotrophic Lateral Sclerosis 1,Adult; Superoxide dismutase [Cu-Zn]

  • ELISA Kit for Superoxide Dismutase 1 (SOD1) Packages (Simulation)
  • ELISA Kit for Superoxide Dismutase 1 (SOD1) Packages (Simulation)
  • ELISA Kit for Superoxide Dismutase 1 (SOD1) Results demonstration
  • SEB960Rb.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the ELISA Kit for Superoxide Dismutase 1 (SOD1)

This assay has high sensitivity and excellent specificity for detection of Superoxide Dismutase 1 (SOD1).
No significant cross-reactivity or interference between Superoxide Dismutase 1 (SOD1) and analogues was observed.

Recovery of the ELISA Kit for Superoxide Dismutase 1 (SOD1)

Matrices listed below were spiked with certain level of recombinant Superoxide Dismutase 1 (SOD1) and the recovery rates were calculated by comparing the measured value to the expected amount of Superoxide Dismutase 1 (SOD1) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 78-93 89
EDTA plasma(n=5) 91-105 98
heparin plasma(n=5) 98-105 102

Precision of the ELISA Kit for Superoxide Dismutase 1 (SOD1)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Superoxide Dismutase 1 (SOD1) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Superoxide Dismutase 1 (SOD1) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the ELISA Kit for Superoxide Dismutase 1 (SOD1)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Superoxide Dismutase 1 (SOD1) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 99-105% 93-105% 84-102% 91-105%
EDTA plasma(n=5) 93-101% 82-99% 95-102% 89-96%
heparin plasma(n=5) 98-105% 95-102% 96-105% 79-92%

Stability of the ELISA Kit for Superoxide Dismutase 1 (SOD1)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the ELISA Kit for Superoxide Dismutase 1 (SOD1)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

Test principle of the ELISA Kit for Superoxide Dismutase 1 (SOD1)

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Superoxide Dismutase 1 (SOD1). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Superoxide Dismutase 1 (SOD1). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Superoxide Dismutase 1 (SOD1), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Superoxide Dismutase 1 (SOD1) in the samples is then determined by comparing the O.D. of the samples to the standard curve.

Related products

Catalog No. Organism species: Oryctolagus cuniculus (Rabbit) Applications (RESEARCH USE ONLY!)
APB960Rb01 Active Superoxide Dismutase 1 (SOD1) Cell culture; Activity Assays.
RPB960Rb01 Recombinant Superoxide Dismutase 1 (SOD1) Positive Control; Immunogen; SDS-PAGE; WB.
PAB960Rb01 Polyclonal Antibody to Superoxide Dismutase 1 (SOD1) WB; IHC; ICC; IP.
PAB960Rb51 Polyclonal Antibody to Superoxide Dismutase 1 (SOD1) WB,IHC
LAB960Rb71 Biotin-Linked Polyclonal Antibody to Superoxide Dismutase 1 (SOD1) WB; IHC; ICC.