ELISA Kit for Amylin

IAPP; IAP; DAP; Islet Amyloid Polypeptide; Insulinoma Amyloid Polypeptide; Diabetes-associated peptide

  • ELISA Kit for Amylin Packages (Simulation)
  • ELISA Kit for Amylin Packages (Simulation)
  • ELISA Kit for Amylin Results demonstration
  • CEA812Ra.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the ELISA Kit for Amylin

This assay has high sensitivity and excellent specificity for detection of Amylin.
No significant cross-reactivity or interference between Amylin and analogues was observed.

Recovery of the ELISA Kit for Amylin

Matrices listed below were spiked with certain level of recombinant Amylin and the recovery rates were calculated by comparing the measured value to the expected amount of Amylin in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 81-102 96
EDTA plasma(n=5) 98-105 102
heparin plasma(n=5) 95-103 98

Precision of the ELISA Kit for Amylin

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Amylin were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Amylin were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the ELISA Kit for Amylin

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Amylin and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 90-98% 80-97% 95-102% 96-103%
EDTA plasma(n=5) 82-101% 95-105% 84-101% 79-104%
heparin plasma(n=5) 87-101% 88-96% 92-101% 82-101%

Stability of the ELISA Kit for Amylin

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the ELISA Kit for Amylin

1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
    And then add 50µL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
7. Add 50µL Stop Solution. Read at 450 nm immediately.

Test principle of the ELISA Kit for Amylin

This assay employs the competitive inhibition enzyme immunoassay technique. A monoclonal antibody specific to Amylin has been pre-coated onto a microplate. A competitive inhibition reaction is launched between biotin labeled Amylin and unlabeled Amylin (Standards or samples) with the pre-coated antibody specific to Amylin. After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of Amylin in the sample. After addition of the substrate solution, the intensity of color developed is reverse proportional to the concentration of Amylin in the sample.

Related products

Catalog No. Organism species: Rattus norvegicus (Rat) Applications (RESEARCH USE ONLY!)
RPA812Ra01 Recombinant Amylin Positive Control; Immunogen; SDS-PAGE; WB.
CPA812Ra11 BSA Conjugated Amylin Immunogen; SDS-PAGE; WB.
CPA812Ra21 OVA Conjugated Amylin Immunogen; SDS-PAGE; WB.
PAA812Ra01 Polyclonal Antibody to Amylin WB; IHC; ICC; IP.
PAA812Ra08 Polyclonal Antibody to Amylin WB; IHC; ICC; IP.
MAA812Ra21 Monoclonal Antibody to Amylin WB; IHC; ICC; IP.
CEA812Ra ELISA Kit for Amylin Enzyme-linked immunosorbent assay for Antigen Detection.
LMA812Ra Multiplex Assay Kit for Amylin ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSA812Ra11 ELISA Kit DIY Materials for Amylin Main materials for "Do It (ELISA Kit) Yourself".
GUESTBOOK