ELISA Kit for Annexin A1 (ANXA1)

ANX-A1; ANX1; LPC1; Lipocortin I; Chromobindin-9; Calpactin II; Phospholipase A2 inhibitory protein

  • ELISA Kit for Annexin A1 (ANXA1) Packages (Simulation)
  • ELISA Kit for Annexin A1 (ANXA1) Packages (Simulation)
  • ELISA Kit for Annexin A1 (ANXA1) Results demonstration
  • SEE787Ra.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the ELISA Kit for Annexin A1 (ANXA1)

This assay has high sensitivity and excellent specificity for detection of Annexin A1 (ANXA1).
No significant cross-reactivity or interference between Annexin A1 (ANXA1) and analogues was observed.

Recovery of the ELISA Kit for Annexin A1 (ANXA1)

Matrices listed below were spiked with certain level of recombinant Annexin A1 (ANXA1) and the recovery rates were calculated by comparing the measured value to the expected amount of Annexin A1 (ANXA1) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 88-99 95
EDTA plasma(n=5) 80-95 88
heparin plasma(n=5) 91-101 97

Precision of the ELISA Kit for Annexin A1 (ANXA1)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Annexin A1 (ANXA1) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Annexin A1 (ANXA1) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the ELISA Kit for Annexin A1 (ANXA1)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Annexin A1 (ANXA1) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 83-101% 81-102% 98-105% 89-96%
EDTA plasma(n=5) 97-104% 92-101% 79-94% 89-97%
heparin plasma(n=5) 78-102% 90-97% 87-101% 79-96%

Stability of the ELISA Kit for Annexin A1 (ANXA1)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the ELISA Kit for Annexin A1 (ANXA1)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

Test principle of the ELISA Kit for Annexin A1 (ANXA1)

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Annexin A1 (ANXA1). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Annexin A1 (ANXA1). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Annexin A1 (ANXA1), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Annexin A1 (ANXA1) in the samples is then determined by comparing the O.D. of the samples to the standard curve.

Related products

Catalog No. Organism species: Rattus norvegicus (Rat) Applications (RESEARCH USE ONLY!)
RPE787Ra02 Recombinant Annexin A1 (ANXA1) Positive Control; Immunogen; SDS-PAGE; WB.
RPE787Ra01 Recombinant Annexin A1 (ANXA1) Positive Control; Immunogen; SDS-PAGE; WB.
APE787Ra01 Active Annexin A1 (ANXA1) Cell culture; Activity Assays.
PAE787Ra01 Polyclonal Antibody to Annexin A1 (ANXA1) WB; IHC; ICC; IP.
LAE787Ra71 Biotin-Linked Polyclonal Antibody to Annexin A1 (ANXA1) WB; IHC; ICC.
MAE787Ra27 Monoclonal Antibody to Annexin A1 (ANXA1) WB; IHC; ICC; IP.
MAE787Ra21 Monoclonal Antibody to Annexin A1 (ANXA1) WB,IHC
MAE787Ra22 Monoclonal Antibody to Annexin A1 (ANXA1) WB; IHC; ICC; IP.
MAE787Ra23 Monoclonal Antibody to Annexin A1 (ANXA1) WB; IHC; ICC; IP.
MAE787Ra24 Monoclonal Antibody to Annexin A1 (ANXA1) WB; IHC; ICC; IP.
MAE787Ra25 Monoclonal Antibody to Annexin A1 (ANXA1) WB; IHC; ICC; IP.
FAE787Ra04 Anti-Annexin A1 (ANXA1) Monoclonal Antibody Flow cytometry.
FAE787Ra02 Anti-Annexin A1 (ANXA1) Monoclonal Antibody Flow cytometry.