ELISA Kit for 5-Hydroxyindoleacetic Acid (5-HIAA)

5HIAA

  • ELISA Kit for 5-Hydroxyindoleacetic Acid (5-HIAA) Packages (Simulation)
  • ELISA Kit for 5-Hydroxyindoleacetic Acid (5-HIAA) Packages (Simulation)
  • ELISA Kit for 5-Hydroxyindoleacetic Acid (5-HIAA) Results demonstration
  • CEB005Ge.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the ELISA Kit for 5-Hydroxyindoleacetic Acid (5-HIAA)

This assay has high sensitivity and excellent specificity for detection of 5-Hydroxyindoleacetic Acid (5-HIAA).
No significant cross-reactivity or interference between 5-Hydroxyindoleacetic Acid (5-HIAA) and analogues was observed.

Recovery of the ELISA Kit for 5-Hydroxyindoleacetic Acid (5-HIAA)

Matrices listed below were spiked with certain level of 5-Hydroxyindoleacetic Acid (5-HIAA) and the recovery rates were calculated by comparing the measured value to the expected amount of 5-Hydroxyindoleacetic Acid (5-HIAA) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 96-104 101
EDTA plasma(n=5) 85-92 88
heparin plasma(n=5) 95-102 98

Precision of the ELISA Kit for 5-Hydroxyindoleacetic Acid (5-HIAA)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level 5-Hydroxyindoleacetic Acid (5-HIAA) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level 5-Hydroxyindoleacetic Acid (5-HIAA) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the ELISA Kit for 5-Hydroxyindoleacetic Acid (5-HIAA)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of 5-Hydroxyindoleacetic Acid (5-HIAA) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 78-104% 82-90% 95-103% 90-99%
EDTA plasma(n=5) 88-103% 81-99% 97-104% 88-103%
heparin plasma(n=5) 82-95% 83-94% 96-104% 98-105%

Stability of the ELISA Kit for 5-Hydroxyindoleacetic Acid (5-HIAA)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the ELISA Kit for 5-Hydroxyindoleacetic Acid (5-HIAA)

1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
    And then add 50µL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
7. Add 50µL Stop Solution. Read at 450 nm immediately.

Test principle of the ELISA Kit for 5-Hydroxyindoleacetic Acid (5-HIAA)

This assay employs the competitive inhibition enzyme immunoassay technique. A monoclonal antibody specific to 5-Hydroxyindoleacetic Acid (5-HIAA) has been pre-coated onto a microplate. A competitive inhibition reaction is launched between biotin labeled 5-Hydroxyindoleacetic Acid (5-HIAA) and unlabeled 5-Hydroxyindoleacetic Acid (5-HIAA) (Standards or samples) with the pre-coated antibody specific to 5-Hydroxyindoleacetic Acid (5-HIAA). After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of 5-Hydroxyindoleacetic Acid (5-HIAA) in the sample. After addition of the substrate solution, the intensity of color developed is reverse proportional to the concentration of 5-Hydroxyindoleacetic Acid (5-HIAA) in the sample.

Related products

Catalog No. Organism species: Pan-species (General) Applications (RESEARCH USE ONLY!)
CPB005Ge21 OVA Conjugated 5-Hydroxyindoleacetic Acid (5-HIAA) Immunogen; SDS-PAGE; WB.
CPB005Ge11 BSA Conjugated 5-Hydroxyindoleacetic Acid (5-HIAA) Immunogen; SDS-PAGE; WB.
PAB005Ge01 Polyclonal Antibody to 5-Hydroxyindoleacetic Acid (5-HIAA) ELISA, CLIA. / IHC-Fr, ICC, IP (predicted).
CEB005Ge ELISA Kit for 5-Hydroxyindoleacetic Acid (5-HIAA) Enzyme-linked immunosorbent assay for Antigen Detection.
LMB005Ge Multiplex Assay Kit for 5-Hydroxyindoleacetic Acid (5-HIAA) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSB005Ge11 ELISA Kit DIY Materials for 5-Hydroxyindoleacetic Acid (5-HIAA) Main materials for "Do It (ELISA Kit) Yourself".
GUESTBOOK