CLIA Kit for Renin (REN)

angiotensinogenase

  • CLIA Kit for Renin (REN) Packages (Simulation)
  • CLIA Kit for Renin (REN) Packages (Simulation)
  • CLIA Kit for Renin (REN) Results demonstration
  • SCA889Hu.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the CLIA Kit for Renin (REN)

This assay has high sensitivity and excellent specificity for detection of Renin (REN).
No significant cross-reactivity or interference between Renin (REN) and analogues was observed.

Recovery of the CLIA Kit for Renin (REN)

Matrices listed below were spiked with certain level of recombinant Renin (REN) and the recovery rates were calculated by comparing the measured value to the expected amount of Renin (REN) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 91-99 95
EDTA plasma(n=5) 88-104 97
heparin plasma(n=5) 79-96 91

Precision of the CLIA Kit for Renin (REN)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Renin (REN) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Renin (REN) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the CLIA Kit for Renin (REN)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Renin (REN) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 88-103% 78-94% 82-105% 95-104%
EDTA plasma(n=5) 88-95% 94-105% 97-104% 79-93%
heparin plasma(n=5) 80-99% 97-105% 91-101% 84-99%

Stability of the CLIA Kit for Renin (REN)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the CLIA Kit for Renin (REN)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

Test principle of the CLIA Kit for Renin (REN)

The microplate provided in this kit has been pre-coated with an antibody specific to Renin (REN). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Renin (REN). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Renin (REN) level in the sample or standard.;

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
APA889Hu61 Active Renin (REN) Cell culture; Activity Assays.
EPA889Hu61 Eukaryotic Renin (REN) Positive Control; Immunogen; SDS-PAGE; WB.
RPA889Hu01 Recombinant Renin (REN) Positive Control; Immunogen; SDS-PAGE; WB.
PAA889Hu01 Polyclonal Antibody to Renin (REN) WB; IHC; ICC; IP.
MAA889Hu23 Monoclonal Antibody to Renin (REN) WB; IHC; ICC; IP.
MAA889Hu21 Monoclonal Antibody to Renin (REN) WB; IHC; ICC; IP.
MAA889Hu25 Monoclonal Antibody to Renin (REN) WB; IHC; ICC; IP.
MAA889Hu24 Monoclonal Antibody to Renin (REN) WB; IHC; ICC; IP.
MAA889Hu22 Monoclonal Antibody to Renin (REN) WB; IHC; ICC; IP.
SEA889Hu ELISA Kit for Renin (REN) Enzyme-linked immunosorbent assay for Antigen Detection.
SCA889Hu CLIA Kit for Renin (REN) Chemiluminescent immunoassay for Antigen Detection.
LMA889Hu Multiplex Assay Kit for Renin (REN) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.