CLIA Kit for Cystatin C (Cys-C)

CST3; Cystatin 3; Gamma Trace; Post-Gamma-Globulin; Neuroendocrine Basic Polypeptide; Amyloid Angiopathy And Cerebral Hemorrhage

  • CLIA Kit for Cystatin C (Cys-C) Packages (Simulation)
  • CLIA Kit for Cystatin C (Cys-C) Packages (Simulation)
  • CLIA Kit for Cystatin C (Cys-C) Results demonstration
  • SCA896Hu.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the CLIA Kit for Cystatin C (Cys-C)

This assay has high sensitivity and excellent specificity for detection of Cystatin C (Cys-C).
No significant cross-reactivity or interference between Cystatin C (Cys-C) and analogues was observed.

Recovery of the CLIA Kit for Cystatin C (Cys-C)

Matrices listed below were spiked with certain level of recombinant Cystatin C (Cys-C) and the recovery rates were calculated by comparing the measured value to the expected amount of Cystatin C (Cys-C) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 84-91 88
EDTA plasma(n=5) 96-105 101
heparin plasma(n=5) 99-105 102

Precision of the CLIA Kit for Cystatin C (Cys-C)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Cystatin C (Cys-C) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Cystatin C (Cys-C) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the CLIA Kit for Cystatin C (Cys-C)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Cystatin C (Cys-C) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 84-98% 94-102% 79-90% 96-103%
EDTA plasma(n=5) 88-101% 78-95% 80-96% 80-101%
heparin plasma(n=5) 78-91% 81-105% 92-103% 90-99%

Stability of the CLIA Kit for Cystatin C (Cys-C)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the CLIA Kit for Cystatin C (Cys-C)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

Test principle of the CLIA Kit for Cystatin C (Cys-C)

The microplate provided in this kit has been pre-coated with an antibody specific to Cystatin C (Cys-C). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Cystatin C (Cys-C). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Cystatin C (Cys-C) level in the sample or standard.;

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
RPA896Hu01 Recombinant Cystatin C (Cys-C) Positive Control; Immunogen; SDS-PAGE; WB.
APA896Hu61 Active Cystatin C (Cys-C) Cell culture; Activity Assays.
EPA896Hu61 Eukaryotic Cystatin C (Cys-C) Positive Control; Immunogen; SDS-PAGE; WB.
PAA896Hu01 Polyclonal Antibody to Cystatin C (Cys-C) WB; IHC; ICC; IP.
LAA896Hu71 Biotin-Linked Polyclonal Antibody to Cystatin C (Cys-C) WB; IHC; ICC.
MAA896Hu22 Monoclonal Antibody to Cystatin C (Cys-C) WB; IHC
RAA896Hu21 Recombinant Antibody to Cystatin C (Cys-C) WB; IHC; ICC.
RAA896Hu22 Recombinant Antibody to Cystatin C (Cys-C) WB; IHC; ICC.
MAA896Hu24 Monoclonal Antibody to Cystatin C (Cys-C) WB,IHC
MAA896Hu25 Monoclonal Antibody to Cystatin C (Cys-C) WB
MAA896Hu26 Monoclonal Antibody to Cystatin C (Cys-C) WB
MAA896Hu21 Monoclonal Antibody to Cystatin C (Cys-C) WB,IHC
LAA896Hu74 Biotin-Linked Monoclonal Antibody to Cystatin C (Cys-C) WB; IHC; ICC.
LAA896Hu76 Biotin-Linked Monoclonal Antibody to Cystatin C (Cys-C) WB; IHC; ICC.
FAA896Hu54 APC-Linked Anti-Cystatin C (Cys-C) Monoclonal Antibody FCM
SEA896Hu ELISA Kit for Cystatin C (Cys-C) Enzyme-linked immunosorbent assay for Antigen Detection.
SCA896Hu CLIA Kit for Cystatin C (Cys-C) Chemiluminescent immunoassay for Antigen Detection.
LMA896Hu Multiplex Assay Kit for Cystatin C (Cys-C) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
PSA896Hu01 Antibody Pair for Cystatin C (Cys-C) ELISA; CLIA; ELISPOT; Luminex; Immunochromatography and other Immunoassays.
KSA896Hu01 ELISA Kit DIY Materials for Cystatin C (Cys-C) Main materials for "Do It (ELISA Kit) Yourself".
GUESTBOOK