CLIA Kit for Cartilage Intermediate Layer Protein 2 (CILP2)

  • CLIA Kit for Cartilage Intermediate Layer Protein 2 (CILP2) Packages (Simulation)
  • CLIA Kit for Cartilage Intermediate Layer Protein 2 (CILP2) Packages (Simulation)
  • CLIA Kit for Cartilage Intermediate Layer Protein 2 (CILP2) Results demonstration
  • SCJ370Hu.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the CLIA Kit for Cartilage Intermediate Layer Protein 2 (CILP2)

This assay has high sensitivity and excellent specificity for detection of Cartilage Intermediate Layer Protein 2 (CILP2).
No significant cross-reactivity or interference between Cartilage Intermediate Layer Protein 2 (CILP2) and analogues was observed.

Precision of the CLIA Kit for Cartilage Intermediate Layer Protein 2 (CILP2)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Cartilage Intermediate Layer Protein 2 (CILP2) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Cartilage Intermediate Layer Protein 2 (CILP2) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Stability of the CLIA Kit for Cartilage Intermediate Layer Protein 2 (CILP2)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the CLIA Kit for Cartilage Intermediate Layer Protein 2 (CILP2)

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
8. Read RLU value immediately.

Test principle of the CLIA Kit for Cartilage Intermediate Layer Protein 2 (CILP2)

The microplate provided in this kit has been pre-coated with an antibody specific to Cartilage Intermediate Layer Protein 2 (CILP2). Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Cartilage Intermediate Layer Protein 2 (CILP2). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Cartilage Intermediate Layer Protein 2 (CILP2) level in the sample or standard.;

Related products

Catalog No. Organism species: Homo sapiens (Human) Applications (RESEARCH USE ONLY!)
RPJ370Hu01 Recombinant Cartilage Intermediate Layer Protein 2 (CILP2) Positive Control; Immunogen; SDS-PAGE; WB.
PAJ370Hu01 Polyclonal Antibody to Cartilage Intermediate Layer Protein 2 (CILP2) WB; IHC; ICC; IP.
MAJ370Hu21 Monoclonal Antibody to Cartilage Intermediate Layer Protein 2 (CILP2) WB; IHC; ICC; IP.
SEJ370Hu ELISA Kit for Cartilage Intermediate Layer Protein 2 (CILP2) Enzyme-linked immunosorbent assay for Antigen Detection.
SCJ370Hu CLIA Kit for Cartilage Intermediate Layer Protein 2 (CILP2) Chemiluminescent immunoassay for Antigen Detection.
LMJ370Hu Multiplex Assay Kit for Cartilage Intermediate Layer Protein 2 (CILP2) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
GUESTBOOK