CLIA Kit for Albumin (ALB)

Serum albumin

  • CLIA Kit for Albumin (ALB) Packages (Simulation)
  • CLIA Kit for Albumin (ALB) Packages (Simulation)
  • CLIA Kit for Albumin (ALB) Results demonstration
  • CCB028Ra.jpg Typical Standard Curve
  • Certificate ISO9001: 2008, ISO13485: 2003 Registered

Specificity of the CLIA Kit for Albumin (ALB)

This assay has high sensitivity and excellent specificity for detection of Albumin (ALB).
No significant cross-reactivity or interference between Albumin (ALB) and analogues was observed.

Recovery of the CLIA Kit for Albumin (ALB)

Matrices listed below were spiked with certain level of recombinant Albumin (ALB) and the recovery rates were calculated by comparing the measured value to the expected amount of Albumin (ALB) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 80-93 88
EDTA plasma(n=5) 78-96 93
heparin plasma(n=5) 91-104 96

Precision of the CLIA Kit for Albumin (ALB)

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Albumin (ALB) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Albumin (ALB) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity of the CLIA Kit for Albumin (ALB)

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Albumin (ALB) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 80-101% 97-105% 78-94% 87-104%
EDTA plasma(n=5) 99-105% 93-101% 80-89% 94-105%
heparin plasma(n=5) 98-105% 86-93% 94-104% 96-103%

Stability of the CLIA Kit for Albumin (ALB)

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Assay procedure summary of the CLIA Kit for Albumin (ALB)

1. Prepare all reagents, samples and standards;
2. Add 50µL standard or sample to each well.
    And then add 50µL prepared Detection Reagent A immediately.
    Shake and mix. Incubate 1 hour at 37°C;
3. Aspirate and wash 3 times;
4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
5. Aspirate and wash 5 times;
6. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
7. Read RLU value immediately.

Test principle of the CLIA Kit for Albumin (ALB)

The microplate provided in this kit has been pre-coated with a monoclonal antibody specific to Albumin (ALB). A competitive inhibition reaction is launched between biotin labeled Albumin (ALB) and unlabeled Albumin (ALB) (Standards or samples) with the pre-coated antibody specific to Albumin (ALB). After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of Albumin (ALB) in the sample. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is reverse proportional to the Albumin (ALB) level in the sample or standard.

Related products

Catalog No. Organism species: Rattus norvegicus (Rat) Applications (RESEARCH USE ONLY!)
NPB028Ra01 Native Albumin (ALB) Positive Control; Immunogen; SDS-PAGE; WB.
RPB028Ra01 Recombinant Albumin (ALB) Positive Control; Immunogen; SDS-PAGE; WB.
EPB028Ra61 Eukaryotic Albumin (ALB) Positive Control; Immunogen; SDS-PAGE; WB.
PAB028Ra01 Polyclonal Antibody to Albumin (ALB) WB; IHC; ICC; IP.
LAB028Ra81 FITC-Linked Polyclonal Antibody to Albumin (ALB) WB; IHC; ICC; IF.
LAB028Ra71 Biotin-Linked Polyclonal Antibody to Albumin (ALB) WB; IHC; ICC.
MAB028Ra22 Monoclonal Antibody to Albumin (ALB) WB; IHC; ICC; IP.
MAB028Ra21 Monoclonal Antibody to Albumin (ALB) WB; IHC; ICC; IP.
CEB028Ra ELISA Kit for Albumin (ALB) Enzyme-linked immunosorbent assay for Antigen Detection.
CCB028Ra CLIA Kit for Albumin (ALB) Chemiluminescent immunoassay for Antigen Detection.
LMB028Ra Multiplex Assay Kit for Albumin (ALB) ,etc. by FLIA (Flow Luminescence Immunoassay) FLIA Kit for Antigen Detection.
KSB028Ra11 ELISA Kit DIY Materials for Albumin (ALB) Main materials for "Do It (ELISA Kit) Yourself".
GUESTBOOK